The AC Voltage measurement has 5 measurement 
positions on the rotary switch : 400mV, 4V, 40V, 
400V and 750V

To measure AC Voltage, connect the Meter as 
follows :

1) Insert the red test lead into the 

V

W

   

terminal

and the black test lead into the

 COM 

terminal.

2) Set the rotary switch to an appropriate

measurement position in V ~ range.

3) Connect the test leads across with the object

being measured.

The measured value is shown on the display, which 
is effective value of sine wave (mean value response).

 

Caution :

n

If the value of voltage to be measured is
unknown, use the maximum measurement
position (750V) and reduce the range step by
step until a satisfactory reading is obtained.

n

The LCD displays “1“ indicating the existing selected
range is overloaded, it is required to select a 
higher range in order to obtain a correct reading.

n

When AC Voltage measurement has been
completed, disconnect the connection between
the testing leads and the circuit under test.

C. AC CURRENT MEASUREMENT

                              

 

(see figure 5)

Warning :

Never attempt an in - circuit current 
measurement where the open circuit voltage 
between terminals and ground is greater than 
60V DC or 30V rms. If the fuse burns out during 
measurement, the Meter may be damaged or the 
operator himself may be hurt. 

 

!

14

  Caution :

n

If the value of voltage to be measured is
unknown, use the maximum measurement
position (1000V) and reduce the range step by
step until a satisfactory reading is obtained.

n

The LCD displays “1” indicating the existing 
selected range is overloaded, it is required to select 
a higher range in order to obtain a correct reading.

n

In each range, the Meter has an input impedance
of approx. 10M

W

 . This loading effect can cause

measurement errors in high impedance circuits. 
If the circuit impedance is less than or equal to
10k

W

 , the error is negligible (0.1% or less).

n

When DC voltage measurement has been
completed, disconnect the connection between
the testing leads and the circuit under test.

B. AC VOLTAGE MEASUREMENT

 

                                         (see figure 4)

  Warning :

To avoid harm to you or damages to the Meter 
from electric shock, please do not attempt to 
measure voltages higher than 1000V or 750V 
rms although readings may be obtained.

( Figure 4 )

Red

Black

~

V

W

CAP

NPN

PNP °C

m

A

mA

10A

CE

OFF

OFF

Hz

SELECT

RANGE

HOLD

Hz/DUTY

hFE

FUSED

10A/15SEC.

MAX

400mA

1000V MAX

V mA

W

COM

10A

207 MK-1(T)

399.9

E

E B C

~

~

~

~

!

!

!

13

Summary of Contents for 207-MK-1

Page 1: ...ontact infrared Thermometer Thermo Hygrometer Thermo Anemometer Wood Moisture Meter Distance Meter Calibrators Gas Analysers Panel meters Battery Testers Digital Hand Held Temperature Indicators Digit...

Page 2: ...vironment you can easily exceed those ratings The best way to avoid damage is to use a clamp meter or to connect a clamp attachment to your DMM To prevent excess current from flowing through your mete...

Page 3: ...nging Digital Multimeter Model 207 MK1 T hereafter referred to as the Meter is a 3 digits Multimeter with steady operations and highly reliable hand held measuring instrument having different measurem...

Page 4: ...1 A m Range 400 A m 4000 m A 40 mA 400 mA AC 2 5 rdg 5dgts 0 01 A 10 A 4 GENERAL SPECIFICATIONS Display 3 digit LCD Maximum reading 3999 with automatic sign and function annunciators Overrange Highes...

Page 5: ...1 C Accuracy 0 1 rdg 4dgts Resolution 0 01 Hz Range 9 999 Hz FREQUENCY 0 001 MHz 9 999 MHz Overload Protection 250 V Rms Sensitivity 0 7V Rms 6 DC CURRENT Accuracy 1 2 rdg 2dgts AC 2 rgd 3dgts DC 1 5...

Page 6: ...o the following rules n Before using the Meter inspect the case Do not use the Meter if it is damaged or the case or part of the case is removed Look for cracks or missing plastic Pay attention to the...

Page 7: ...nal button operations 10 V W CAP NPN PNP C m A mA 10A CE OFF OFF Hz SELECT RANGE HOLD Hz DUTY hFE FUSED 10A 15SEC MAX 400mA 1000V MAX VW mA COM 10A 207 MK 1 T 399 9 E E B C The Multimeter Structure se...

Page 8: ...display is not on otherwise false readings may be provided n Pay extra attention to the symbol which is located besides the input terminals of the Meter before carrying out measurement A DC VOLTAGE ME...

Page 9: ...uit voltage between terminals and ground is greater than 60V DC or 30V rms If the fuse burns out during measurement the Meter may be damaged or the operator himself may be hurt 14 Caution n If the val...

Page 10: ...ion and range for the measurement When the testing leads are connected to the current terminals do not parallel them across any circuit The AC current measurement has 5 measurement positions on the ro...

Page 11: ...range for the measurement When the testing leads are connected to the current terminals do not parallel them across any circuit The DC current measurement has 5 measurement positions on the rotary sw...

Page 12: ...10A 15SEC MAX 400mA 1000V MAX V mA W COM 10A 207 MK 1 T 399 9 E E B C Capacitor 20 Note n The test leads can add 0 1W to 0 3W of error to the Resistance measurement To obtain precision readings in lo...

Page 13: ...DUTY hFE FUSED 10A 15SEC MAX 400mA 1000V MAX V mA W COM 10A 207 MK 1 T 399 9 E E B C 22 Caution n Never connect high voltage to the input sockets with the switch in Capacitance range n Using Capacitan...

Page 14: ...value Figure 11 Red Black V W CAP NPN PNP C m A mA 10A CE OFF OFF Hz SELECT RANGE HOLD Hz DUTY hFE FUSED 10A 15SEC MAX 400mA 1000V MAX V mA W COM 10A 207 MK 1 T 399 9 E E B C 24 The frequency measure...

Page 15: ...ads and the circuit under test Testing for Continuity To test for continuity connect the Meter as below 1 Insert the red test lead into VW terminal and the black test lead into the COM terminal 2 Set...

Page 16: ...thorized dealer This warranty does not apply for damaged Ic s fuses burnt PCB s disposable batteries carrying case test leads or to any product which in KUSAM MECO s opinion has been misused altered n...

Reviews: