Accuracy  is  ± (% reading   number  of  digits)  or  otherwise 

specified, at 23°C ± 5°C & less than 75% R.H.

DC VOLTAGE

Range

Resolution

Accuracy

400.0  mV

4.000     V

40.00     V

400.0     V

600        V

100.0  

m

V

1        mV

10      mV

100    mV

1           V

± (0.3% rdg + 4 dgts)

± (0.5% rdg + 3 dgts)

± (0.5% rdg + 3 dgts)

± (0.5% rdg + 3 dgts)
± (1.0% rdg + 4 dgts)

NMRR :

 >50dB @ 50/60Hz

CMRR :

 >120dB @ DC, 50/60Hz, Rs = 1k

W

Input Impedance :

 10M

W

, 30pF nominal (1000M

W

 for 400.0mV range)

AC VOLTAGE (50Hz ~ 500Hz)

Range

Resolution

Accuracy

4.000     V

40.00     V

400.0     V

600        V

1        mV

10      mV

100    mV

1           V

± (1.5% rdg + 5 dgts)

± (2.0% rdg + 5 dgts)

CMRR :

 >60dB @ DC to 60Hz, Rs = 1k

W

Input Impedance :

 10M

W

, 30pF nominal 

± (1.5% rdg + 5 dgts)

± (1.5% rdg + 5 dgts)

True  RMS  accuracy  of  ACV is specified from 5% to 100% of range, &

ACA  is  specified  from  10%  to  100%  of  range,  or  otherwise  specified.

Maximum  Crest  Factor  <1.75  :  1  at  full  scale  &  <3.5  :  1  at  half  scale,

and  with  frequency  spectrums,  besides  fundamentals,  fall  within  the 

meter specified AC bandwidth for non-sinusoidal waveforms.

Summary of Contents for KM 111M

Page 1: ...Tester Digital Panel Meters DC Power Supplies High Voltage Detector Calibrators Gas Analysers Frequency Counter Function Generator Phasing Sticks Battery Tester Waterproof Pen Testers Solar Power Mete...

Page 2: ...nt Even if you correctly insert your clamp meter in to the circuit you can still damage your meter Don t try to measure current in excess of your meter s capacity Check the current capacity of the Cla...

Page 3: ...c Per IEC61010 Overvoltage Installation Category Overvoltage Category II Equipment of is energy consuming equipment to be supplied from the fixed installation Overvoltage Category II Note Examples inc...

Page 4: ...7 7 3 4 Average KM 111M True RMS KM112M...

Page 5: ...nal 1000MW for 400 0mV range AC VOLTAGE 50Hz 500Hz Range Resolution Accuracy 4 000 V 40 00 V 400 0 V 600 V 1 mV 10 mV 100 mV 1 V 1 5 rdg 5 dgts 2 0 rdg 5 dgts CMRR 60dB DC to 60Hz Rs 1kW Input Impedan...

Page 6: ...2 0 rdg 4 dgts 1 0 rdg 4 dgts MAX HOLD Specified accuracy 50 digits for changes 25ms in duration AUDIBLE CONTINUITY TESTER Audible threshold between 5 and 120 WW DIODE TESTER Open Circuit Voltage 1 6...

Page 7: ...9 10 CAUTION Before and after hazardous voltage measurements test the voltage function on a known source such as line voltage to determine proper meter functioning...

Page 8: ...11 12...

Page 9: ...13 14...

Page 10: ...15 16...

Page 11: ...17 16 18...

Page 12: ...hority to extend a greater or different warranty on behalf of KUSAM MECO KUSAM MECO s warranty obligation is limited at option free of charge repair or replacement of a defective product which is retu...

Reviews: