background image

18

®

 

Application notes for flame sensors:

DC

A  function  is  useful  for  HVAC/R  flame  sensor  applications. 

The  0.1

A  resolution  can  identify  the  minute  current  changes  in 

flame  detector  applications.  Flame  signal  current  check  should
indicate steady flame signal of at least 2

A for a rectification type, 

or 1.5

A for an ultraviolet type (8

A for self checking systems). If a 

flame signal current with inadequate strength or fluctuation beyond 
10%, check the following to avoid the risk of unwanted flame relay 
dropout :

CAUTION

Discharge capacitors before making any measurement. Large 
value capacitors should be discharged through an appropriate 
resistance load.

A, mA and A Current

Press SELECT button momentarily to toggle between DC and AC. 
Last  selection  will  be  saved  as  power  up  default  for  repeat 
measurement convenience.

®

17

TM

Resistance, BeepLit  Continuity”;

Press the SELECT button momentarily to toggle the functions. Last
selection will be saved as power up default for repeat measurement 
convenience.

 

 BeepLitTM Continuity 

function is having improved convenience 

for  checking  wiring  connections  and  operation  of  switches.  A 
continuous  beep  tone  together  with  flashing  display  backlight 
indicate  a  complete  wire.  Such  audible  and  visible  indications 
improve continuity readabilities in noisy working environments.

Summary of Contents for KM 233

Page 1: ...Digital Lux Meter Network Cable Tester Power Factor Regulator Earth Resistance Tester Digital Panel Meters DC Power Supplies High Voltage Detector Calibrators Gas Analysers Frequency Counter Function...

Page 2: ...TABLE OF CONTENTS TITLE PAGE NO I Safety II Cenelec Directives III Special Features IV General Specifications V Electrical Specifications VI Product Description VII Maintenance VIII Test Certificate I...

Page 3: ...4mm for CAT III CAT IV ratings Refer to the category markings on your probe assemblies as well as on the add on accessories like detachable Caps or Alligator Clips if any for applicable rating changes...

Page 4: ...004 108 CU and RoHS directive 2011 65 EU 04 IV GENERAL SPECIFICATIONS Sensing True RMS sensing Display 3 5 6 digits 6 000 counts Update Rate 5 per second nominal o o Operating Temperature 10 C to 45 C...

Page 5: ...User s manual Optional Purchase Accessories BMH 01 magnetic hanger BMP 25x soft carrying pouch ELECTRICAL SPECIFICATIONS Accuracy is given as of reading digits number of digits or o o otherwise specif...

Page 6: ...0 3 rdg 2dgts 0 4 rdg 2dgts 0 4 rdg 2dgts 0 2 rdg 2dgts 60 00mV 600 0V 1000V 60 00V 600 0mV 6 000V Range Resolution 10 V 100 V 10 V 100 V 10 V 100 V 1 mV 10 mV 100 mV 1 V 08 BeepLit Continuity Tester...

Page 7: ...00M 100 mV 1 V 100 mV 1 V 100 m 1 10 100 1 k 10 k 10 DC Current AC Current Range Range Burden Voltage Burden Voltage Accuracy Accuracy 600 0 A 600 0 A 6000 A 6000 A 60 00mA 60 00mA 600 0mA 600 0mA 6...

Page 8: ...nput Jack for 10A 20A for 30sec current function 5 Input Jack for all functions EXCEPT A mA A current functions 6 Common Ground reference Input Jack for all functions 7 Input Jack for A and mA current...

Page 9: ...oning AutoV LoZ mode AutoV automatically selects measurement function of DCV or ACV based on their input levels via the test leads The input also provides a low ramp up impedance LoZ to drain ghost vo...

Page 10: ...t is recommended to first measure the signal voltage or current level and activate Hz function in that range to get the most appropriate trigger level If the Hz reading becomes unstable select lower s...

Page 11: ...ny measurement Large value capacitors should be discharged through an appropriate resistance load A mA and A Current Press SELECT button momentarily to toggle between DC and AC Last selection will be...

Page 12: ...e over 165 oF 74 oC at photocell 19 CAUTION Using resistance and continuity function in a live circuit will produce false results and may damage the instrument In many cases the suspected component mu...

Page 13: ...feature 21 Electric Field EF Detection Press the EF button momentarily to toggle EF Detection feature The meter displays EF H when it is ready If it is too sensitive for your applications press Level...

Page 14: ...n zero readings for Hz function 5 Electric field signal present for EF function 6 Significant movement indication as in Phase Rotation functions 23 Relative Zero mode Relative Zero allows the user to...

Page 15: ...2 for A current input 11A 1000V DC AC IR 20kA F fuse or better Dimension 10 x 38mm 26 MUMBAI TEST CERTIFICATE TRMS DIGITAL MULTIMETER This Test Certificate warrantees that the product has been inspect...

Page 16: ...aminated or damaged by accident or abnormal conditions of operation or handling KUSAM MECO authorized dealer shall extend this warranty on new and unused products to end user customers only but have n...

Reviews: