background image

®

05

     Ohm function :

 

          Total Accuracy = Specified Ac 15 digits

     Other functions :

          Total Accuracy = Specified Accuracy

Performance above 3V/m is not specified

  

Overload Protection :

 

        A

 

& mA :

 0.4A/1000V DC/AC rms, IR 30kA, F fuse or better

       

A :

 11A/1000V DC/AC rms, IR 20kA, F fuse or better

       

V & AutoV :

 1100V DC/AC rms

       

mV, Ohm & others :

 1000V DC/AC rms

  

Low Battery :

 Below approx. 2.5V

  

Power Supply :

 1.5V AAA Size battery X 2

  

Power Consumption (typical) :

 3.2mA

  

APO Consumption (typical) : 

10

  

APO Timing :

 Idle for 30 minutes

  

Dimension :

 161*80*50mm L*W*H (With Holster) 

  

Weight :

 Approx. 334 gm (With Holster)

  

Accessories :

 Test lead pair; Batteries installed; User's manual.

  

Optional Purchase Accessories :

  BMH-01 magnetic hanger; 

     BMP-25x soft carrying pouch.

®

ELECTRICAL SPECIFICATIONS :

Accuracy  is  given  as 

(%  of  reading  digits  +  number  of  digits)  or 

o

o

otherwise  specified  @  23 C 

  5 C  ACV  &  ACA  accuracies  are 

specified  from  1  %  to  100  %  of  range  or  otherwise  specified. 
Maximum  Crest  Factor  <2:1  at  full  scale  &  <4:1  at  half  scale,  and 
with frequency components fall within the meter specified frequency 
bandwidth for non-sinusoidal waveforms 

AC VOLTAGE

Input Impedance:

 10M

, 54pF nominal

1) 

<5d non-zero residue may appear when backlight is on, which 

will not affect the specified measuring range and accuracy

Accuracy

50Hz ~ 60Hz

600.0V

±(0.7%rdg+3dgts)

±(2.0%rdg+3dgts)

1)

6.000V

1000 V

60.00V

45Hz ~ 440kHz

600.0V

60.00V

1000 V

1)

6.000V

Range

Resolution

1     mV

10   mV

100 mV

1        V

1     mV

10   mV

100 mV

1        V

06

Summary of Contents for KM 233

Page 1: ...Digital Lux Meter Network Cable Tester Power Factor Regulator Earth Resistance Tester Digital Panel Meters DC Power Supplies High Voltage Detector Calibrators Gas Analysers Frequency Counter Function...

Page 2: ...TABLE OF CONTENTS TITLE PAGE NO I Safety II Cenelec Directives III Special Features IV General Specifications V Electrical Specifications VI Product Description VII Maintenance VIII Test Certificate I...

Page 3: ...4mm for CAT III CAT IV ratings Refer to the category markings on your probe assemblies as well as on the add on accessories like detachable Caps or Alligator Clips if any for applicable rating changes...

Page 4: ...004 108 CU and RoHS directive 2011 65 EU 04 IV GENERAL SPECIFICATIONS Sensing True RMS sensing Display 3 5 6 digits 6 000 counts Update Rate 5 per second nominal o o Operating Temperature 10 C to 45 C...

Page 5: ...User s manual Optional Purchase Accessories BMH 01 magnetic hanger BMP 25x soft carrying pouch ELECTRICAL SPECIFICATIONS Accuracy is given as of reading digits number of digits or o o otherwise specif...

Page 6: ...0 3 rdg 2dgts 0 4 rdg 2dgts 0 4 rdg 2dgts 0 2 rdg 2dgts 60 00mV 600 0V 1000V 60 00V 600 0mV 6 000V Range Resolution 10 V 100 V 10 V 100 V 10 V 100 V 1 mV 10 mV 100 mV 1 V 08 BeepLit Continuity Tester...

Page 7: ...00M 100 mV 1 V 100 mV 1 V 100 m 1 10 100 1 k 10 k 10 DC Current AC Current Range Range Burden Voltage Burden Voltage Accuracy Accuracy 600 0 A 600 0 A 6000 A 6000 A 60 00mA 60 00mA 600 0mA 600 0mA 6...

Page 8: ...nput Jack for 10A 20A for 30sec current function 5 Input Jack for all functions EXCEPT A mA A current functions 6 Common Ground reference Input Jack for all functions 7 Input Jack for A and mA current...

Page 9: ...oning AutoV LoZ mode AutoV automatically selects measurement function of DCV or ACV based on their input levels via the test leads The input also provides a low ramp up impedance LoZ to drain ghost vo...

Page 10: ...t is recommended to first measure the signal voltage or current level and activate Hz function in that range to get the most appropriate trigger level If the Hz reading becomes unstable select lower s...

Page 11: ...ny measurement Large value capacitors should be discharged through an appropriate resistance load A mA and A Current Press SELECT button momentarily to toggle between DC and AC Last selection will be...

Page 12: ...e over 165 oF 74 oC at photocell 19 CAUTION Using resistance and continuity function in a live circuit will produce false results and may damage the instrument In many cases the suspected component mu...

Page 13: ...feature 21 Electric Field EF Detection Press the EF button momentarily to toggle EF Detection feature The meter displays EF H when it is ready If it is too sensitive for your applications press Level...

Page 14: ...n zero readings for Hz function 5 Electric field signal present for EF function 6 Significant movement indication as in Phase Rotation functions 23 Relative Zero mode Relative Zero allows the user to...

Page 15: ...2 for A current input 11A 1000V DC AC IR 20kA F fuse or better Dimension 10 x 38mm 26 MUMBAI TEST CERTIFICATE TRMS DIGITAL MULTIMETER This Test Certificate warrantees that the product has been inspect...

Page 16: ...aminated or damaged by accident or abnormal conditions of operation or handling KUSAM MECO authorized dealer shall extend this warranty on new and unused products to end user customers only but have n...

Reviews: