- 10 -
NON-CONTACT EF-DETECTION
Typical Voltage
20V (tolerance : 10V ~ 36V)
55V (tolerance : 23V ~ 85V)
110V (tolerance : 59V ~ 600V)
Bar-Graph
Indication
-
- - -
- - - - -
Indication :
Bar-graph segments & audible beep
tones proportional to the field strength
Detection Frequency :
50/60Hz
Detection Antenna :
Top side of the stationary jaw
Probe-Contact EF-Detection: For more precise
indication of live wires, such as distinguishing
between live and ground connections, use the
Red (+) test probe for direct contact measurement.
CAPACITANCE
60.00 nF
±(2.0%rdg + 5dgts)
600.0 nF
6.000
m
F
60.00
m
F
±(3.5%rdg + 5dgts)
2)
Resolution
Range
±(4.0%rdg + 5dgts)
2)
1)
Accuracy
600.0
m
F
2000
m
F
1)
Accuracies with film capacitor or better
2)
Temperature Coefficient : 0.25 x (specified accuracy) / °C
@ (0°C ~ 18°C or 28°C ~ 40°C)
10 pF
100 pF
1 nF
10 nF
100 nF
1
m
F
- 11 -
TM
AUTOCHECK _DCV
Accuracy
Resolution
Range
TM
AutoCheck Lo-Z DCV Threshold :
>+1.5VDC & <-1.5VDC nominal.
TM
AutoCheck Lo-Z DCV input impedance :
Initially approx. 2.5k
W
, 600pF nominal; impedance increases abruptly within a
fraction of a second as display voltage is above 50V (typical).
Ended up impedances vs display voltages typically are :
10 k
W
@ 100V
60 k
W
@ 300V
200k
W
@ 600V
420k
W
@ 1000V
6.000 V
±(1.3%rdg + 5dgts)
60.00 V
600.0 V
1000 V
1 mV
10 mV
100 mV
1 V
TM
OHM & AUTOCHECK _OHM
1)
600.0
W
±(0.5%rdg + 5dgts)
6.000 K
W
60.00 K
W
600.0 K
W
6.000 M
W
Open Circuit Voltage :
0.45VDC typical.
1)
TM
AutoCheck OhmThreshold :
<10.00M
W
nominal.
±(0.8%rdg + 5dgts)
±(1.2%rdg + 5dgts)
Accuracy
Resolution
Range
40.00 M
W
±(2.3%rdg + 5dgts)
0.1
W
1
W
10
W
100
W
1 K
W
10 K
W