For non-invasive current measurements, press the jaw trigger and
clamp the jaws around conductor(s) of only one single pole of a
circuit for load current measurement. Make sure the jaws are
completely closed, or else it will introduce measurement errors.
Enclosing conductor(s) of more than one pole of a circuit may result
in differential current (like identifying leakage current)
measurement. Locate the conductor(s) at the Jaws center as much
as possible to get the best measuring accuracy. For removal, press
the jaw trigger and remove the jaws from the conductor(s).