8. Data setting:
When using the character on the nameplate to input data, the
previous setting will appear, and the place that waits to be entered
will flash. INPUT starts from the largest digit to enter, the largest
digit only 1, In that case the original setting will flash. The original
setting is 0, and the bottom part of seven segments will flash. Press
“1” is 1, press any keys is 0, press “ENTER” to exit without change.
TOL Hi/Lo Limits setting without largest digit. There is no
annunciator flashing After entering 5 digits, In the mean time enters
+/- symbols, press “0” to change +/- symbols, press other keys to
input + symbol.
*Caution: Press “ENTER”, emit two beeps, the data store in
volatile storage. The data will store in nonvolatile storage region
while the meter power off. In this mode the automatic power-off
feather is disabled.
(10) REL RELATIVE MODE (ONLY MAIN DISPLAY)
Press “REL” key to the Relative mode. The displayed reading is
stored as a reference value, the display is zeroed and annunciator “
Δ
”
is displayed. Press REL key again to exit the Relative mode.
For example: the displayed reading is 100.0, then press REL to store
as a reference value, and the display become zero, to store 100.0 as a
standard reference value, if our input signal is 99.5, then the reading
will be 99.5-100.0 that equals -0.5. The user (see “SET” in this
manual can also set the relative value). When the desired relative
value has been entered, press REL key to enter the Relative mode, and
press SET key to use Relative value as a reference value. Press REL
again to exit the relative Mode.
(11) HI/LO LIMITS (ONLY MAIN DISPLAY)
Press “Hi/Lo LIMITS” to enter the Hi/Lo LIMITS mode, and change
to manual range mode, the original Hi/LO LIMITS value and
annunciators “ ” “ ” appear in the same time individually. When the
input exceeds Hi limits, the “ ” is blinking and emits a continuous
tone. When the input goes below the Lo limits, the “ ” is blinking
and the beeper emits a pulse tone. This warns users that out of setting
ranges, press “Hi/Lo LIMITS” again to exit this mode. In additions,
the meter will ignore and not to record if overload (“OL”) situation
happens in the comparative processes or in the capacitance range
≤
50
counts.
(12) TOL
Press “TOL” to enter the tolerance mode, and change to manual range
mode automatically, the original preset standard value and
annunciators “TOL” appear in the same time individually. How to set
standard value please refer to SET in this manual. When entering TOL
mode, the Main display is the present value and the second display is
tolerance value.
There are 4 preset values in the TOL mode for instant use, just to
press TOL again to cycle through 1%, 5%, 10%, 20%, then back to
the present. When entering the TOL mode, and the annunciator “ ”
“ ” will appear in the same time, when the input exceeds Hi limits,
the “ ” is blinking and also emits a continuous tone. When the input
goes below the LO limits, the “ ” is blinking and the beeper emits a
pulse tone. This warns users that out of setting ranges.
The standard value can be set by SET function, please refer to SET
setting in this manual. When standard value have set, and press
“TOL” then press “SET” to use preset TOL setting. In additions, the
meter will ignore and not to record if overload (“OL”) situation
happens in the comparative processes or in the capacitance range
≤
50
counts. Press “TOL” 2 sec to exit this function.