8. Data setting: 

When  using  the  character  on  the  nameplate  to  input  data,  the 
previous setting will appear, and the place that waits to be entered 
will  flash.  INPUT  starts  from  the  largest  digit  to  enter,  the  largest 
digit only 1, In that case the original setting will flash. The original 
setting is 0, and the bottom part of seven segments will flash. Press 
“1” is 1, press any keys is 0, press “ENTER” to exit without change. 
TOL  Hi/Lo  Limits  setting  without  largest  digit.  There  is  no 
annunciator flashing After entering 5 digits, In the mean time enters 
+/-  symbols,  press  “0”  to  change  +/-  symbols,  press  other  keys  to 
input + symbol. 
*Caution:  Press  “ENTER”,  emit  two  beeps,  the  data  store  in 

volatile  storage.  The  data  will  store  in  nonvolatile  storage  region 
while the  meter power off. In this  mode the  automatic power-off 
feather is disabled. 

 

(10) REL RELATIVE MODE (ONLY MAIN DISPLAY) 

Press  “REL”  key  to  the  Relative  mode.  The  displayed  reading  is 
stored as a reference value, the display is zeroed and annunciator “

Δ

” 

is displayed. Press REL key again to exit the Relative mode. 
For example: the  displayed reading  is 100.0, then press REL to store 
as a reference value, and the display become zero, to store 100.0 as a 
standard  reference  value,  if our  input  signal  is 99.5,  then the  reading 
will  be  99.5-100.0  that  equals  -0.5.  The  user  (see  “SET”  in  this 
manual  can  also  set  the  relative  value).  When  the  desired  relative 
value has been entered, press REL key to enter the Relative mode, and 
press SET key to use Relative value as a reference value. Press REL 
again to exit the relative Mode. 
 

 

 

 

 

(11) HI/LO LIMITS (ONLY MAIN DISPLAY) 

Press “Hi/Lo LIMITS” to enter the Hi/Lo LIMITS mode, and change 
to  manual  range  mode,  the  original  Hi/LO  LIMITS  value  and 
annunciators “ ” “ ” appear in the same time individually. When the 
input  exceeds  Hi  limits,  the  “ ”  is  blinking  and  emits  a  continuous 
tone.  When  the  input  goes  below  the  Lo  limits,  the  “ ”  is  blinking 
and the beeper emits a pulse tone. This warns users that out of setting 
ranges,  press  “Hi/Lo  LIMITS”  again  to  exit  this  mode.  In  additions, 
the  meter  will  ignore  and  not  to  record  if  overload  (“OL”)  situation 
happens in the comparative processes or in the capacitance range 

50 

counts. 

 

(12) TOL 

Press “TOL” to enter the tolerance mode, and change to manual range 
mode  automatically,  the  original  preset  standard  value  and 
annunciators “TOL” appear in the same time individually. How to set 
standard value please refer to SET in this manual. When entering TOL 
mode, the Main display is the present value and the second display is 
tolerance value. 

 

There  are  4  preset  values  in  the  TOL  mode  for  instant  use,  just  to 
press  TOL  again  to  cycle  through  1%,  5%,  10%,  20%,  then  back  to 
the  present.  When  entering  the TOL  mode,  and  the  annunciator  “ ” 
“ ”  will  appear  in  the  same  time,  when  the  input  exceeds  Hi  limits, 
the “ ” is blinking and also emits a continuous tone. When the input 
goes below the LO limits, the “ ” is blinking and the beeper emits a 
pulse tone. This warns users that out of setting ranges. 

 

The  standard  value  can  be  set  by  SET  function,  please  refer  to  SET 
setting  in  this  manual.  When  standard  value  have  set,  and  press 
“TOL” then press  “SET” to use preset TOL setting.  In additions, the 
meter  will  ignore  and  not  to  record  if  overload  (“OL”)  situation 
happens in the comparative processes or in the capacitance range 

50 

counts. Press “TOL” 2 sec to exit this function. 

Summary of Contents for KM-459

Page 1: ...nction Generator Phasing Sticks Battery Tester Waterproof Pen Testers Solar Power Meter EMF Detector Wood Paper Grain Moisture Meter Transistorised Electronic Analog Digital Insulation Resistance Testers upto 10 KV Digital Sound Level Meter Sound Level Calibrator Digital contact Non contact Type Tachometer Digital Non contact infrared Thermometer Maximum Demand Controller Digital Power Meter Digit...

Page 2: ...orage environment 20 C to 60 C at 0 to 80 RH with battery removed from meter Power Single standard 9 volt battery NEDA 1604 JIS 006P IEC 6F22 External power Minimum 12V maximum 15V and at least minimum 50mA Auto power As the APO is displayed on the LCD the meter will shut down by itself if unused for about 10 minutes press the power key to resume power on mode The meter will cancel auto power off ...

Page 3: ...2 Communication is activated annunciator R Recording mode annunciator MAX Maximum reading annunciator MIN Minimum reading annunciator AVG Average reading annuciator AUTO Autoranging indicator H Data hold annunciator SET Set mode annunciator Relative mode annunciator TOL Tolerance mode annunciator PAL Parallel mode annunciator SER Series mode annunciator D Dissipation factor annunciator Q Quality f...

Page 4: ... the battery power is weakening Beeper tone indicator for tolerance mode MKΩ umH nd u Ictance Henry annunciator munpF Capacitance ara annunciator SPECIFICATIONS CAPACITANCE Test Frequency 120Hz 10mF 1 F μ 10nF 1nF 10pF 1pF 2000 F μ 200μ F 20μ F 2000nF 200nF 20nF Range Min Max Cx DF Note Resistance O mhannunciator ...

Page 5: ...KHz 2000 F μ 10nF 1nF 10pF 1pF 0 1pF 200 F μ 20μ F 2000nF 200nF 20nF 2000pF Range Min Max Cx DF Note INDUCTANCE Test Frequency 120Hz Range 1H 10000H 1mH 10 H μ 1 H μ 2000H 200H 20H 200mH 20mH Min Max Lx DF 0 5 DF DF 0 5 Note ...

Page 6: ... H μ 1 H μ 2000H 200H 20H 200mH 20mH 2000 H μ Range Min Max Lx DF 0 5 DF DF 0 5 Note RESISTANCE 10MΩ 2MΩ 200KΩ 20KΩ 2KΩ 200Ω 20Ω Note In 20 range effective readings must over 20 counts Range Min Max Test Freq 120Hz Note Test Freq 1KHz ...

Page 7: ...cked and damaged Please check periodically To avoid electric shock discharge the circuits completely before taking any measurements CAUTION When it appears to have abnormal situations such as you cannot turn on the meter to operate 1 It is normal situation when you cannot turn on the meter after turning off the meter just for few seconds Please wait a moment to turn it on again 2 When you cannot o...

Page 8: ...ies equivalent circuit mode SER 4 RANGE BUTTON Press RANGE button to select the Manual Range mode and turn off the AUTO annunciator The meter remains in the range it was in when manual ranging was selected In the Manual Range mode Each time you press RANGE button the range and the input range annunciator increments and a new value is displayed To exit the Manual Range mode and return to autorangin...

Page 9: ... Press Min Max to enter the Min Max Avg mode and stop the auto shut down function In Addition to power and hold key the other keys cannot be activated When the meter samples about 6 times and then beeper will emit a sound When a new Max Min data being recorded and then beeper will emit 2 sounds Main display value second display value cycles through Min Max key present value parameter value Max val...

Page 10: ... HI LO LIMITS ONLY MAIN DISPLAY Press Hi Lo LIMITS to enter the Hi Lo LIMITS mode and change to manual range mode the original Hi LO LIMITS value and annunciators appear in the same time individually When the input exceeds Hi limits the is blinking and emits a continuous tone When the input goes below the Lo limits the is blinking and the beeper emits a pulse tone This warns users that out of sett...

Page 11: ...will sound Replace fuse only with the original type 0 1A 250V fast acting fuse Cleaning Periodically wipe the case with a damp cloth and detergent do not use abrasives or solvents RS 232 COMMAND TABLE Use IR as the interface of Data transmission and use external computer to start RS 232 functions RS 232 Interface Parameter Baud rate 1200 Parity check EVEN Data buts 7 Stop bits 1 1 Setup Selections...

Page 12: ...etting value V 19999 Limits Hi setting value W 19999 Limits Lo setting value X 19999 TOL SET setting value Y 19999 TOL SET Hi setting value Z 19999 TOL SET Lo setting value After receiving of setting data from PC U V W X Y Z and will send out the receiving data again then it is convenient for PC to check the accuracy of data output are total 7 digits c Command BXXXXXX Exit the SETUP mode 2 Read Da...

Page 13: ... OL 28 S SET _ normal 29 F FUSE _ normal 30 H HOLD _ normal 31 R Present value M Maximum value I Minimum value X Max Min value A Average value _ normal 32 R REL S REL SET _ normal 33 L LIMITS _ normal 34 T TOL S TOL SET _ normal 35 B Backlight _ normal 36 A Adapter insert _ normal 37 B Low Battery _ normal 38 CR ASCII 0DH 39 nl LF ASCII 0AH ...

Page 14: ... 22 4 RS 232 Output Chart for Second Display Rs 100Ω R Rs 1KΩ 10ΚΩ R Rs 100KΩ 1 2 3 4 5 5 FORMULA RP RS 1 Q2 CP CS 1 1 D2 CS CP 1 D2 LP LS 1 1 Q2 LS LP Q2 1 Q2 Range Q D R ...