background image

11

"REL"  symbol  appears  on  LCD.  The  secondary 
display will show the percentage of relative value
REL%

The REL% = (present value - initial value) / initial 
value * 100%.

Press 

REL%

  key  again  to  show  the  current 

readings on primary display and the "REL" symbol 
will  be  blinking.  The  percentage  range  is  from 
-99.9% ~ 99.9% when the present value is larger 
than  double  of  initial  value,  the  "OL%"  indication 
will be shown on the secondary display.

During   relative   measurement,   analog  bar  is 
always  indicating    the  present  measurement 
value but not the relative value.

6. Open / Short calibration

1)  Press 

CAL

 key larger than 2 seconds to start the

     open / short calibration procedure.

2)  In open calibration mode, the secondary display
    will show "Open". There are two ways for open 
     state input:

a)   When using square terminals, the square 

 

      and                                                                the air 
      (shown  in figure 4);

terminals

L

CUR / 

L

POT / 

H

POT / 

H

       CUR     

terminals hang in

5. Relative mode

During relative measurement the meter remembers
the  current  readings  on  primary  display  (called
initial value) when pressing the 

REL%

 key, and 

12

b)   When using                                                insert
      the black and red testing lines with alligator clip 
      into the "       ","       ",terminal and "
      

                            

      

terminal respectively. Connect the mouth of clips 

      (shown in figure 7).    

L

CUR / 

L

POT / 

H

POT / 

H

CUR   

terminals

L

CUR  

L

POT  

CUR

H     ", "H
  

POT

".

a)   When using square terminals, insert the short 
      soket    to   the   square   terminal   and  make
                                     

 

       
                                                               

  terminals  hang  in the air 

      (shown in figure 6); 

L

CUR / 

L

POT / 

H

POT / 

H

CUR  

b)  When using                                                     insert
     the black and red testing lines with alligator clip into
     the "       ","       ",terminal and "                                 
     respectively. (shown in figure 5).   

H

CUR  

terminals

POT

L

CUR / 

L

POT / 

H

POT / 

L

CUR  

L

POT  

H     ", "H      terminal

CUR

GUARD

GUARD

INPUT

Rx Lx Cx

L

CUR

L

POT

H

POT

H

CUR

GUARD

GUARD

INPUT

Rx Lx Cx

L

CUR

L

POT

H

POT

H

CUR

Figure 4

Figure 5

3)    Press 

CAL

  key  and  the  30-second  countdown 

      will  be  shown  on  LCD  panels.  If  the  open 
      calibration is finished, the PASS or  FAIL symbol   
        will  be  shown  on    the    primary  display.    Press   
       

CAL

  key    again  to  save  the    calibration    data 

      and enter into the  short  calibration mode.

4)  In  short  calibration  mode,  the  secondary 
      display will show "Srt". There are two ways for 
      short state input:

Summary of Contents for KM 520B

Page 1: ...meter Wood Paper Grain Moisture Meter Distance Meter Network Cable Tester Digital Hand Held Temperature Indicators Digital Lux Meter Thermal Imaging Camera Power Factor Regulator Maximum Demand Contro...

Page 2: ...o turn off the instrument when replacing the battery or DC power adapter Although the instrument has a protection for shocks inputing dc voltage or current will still damage the instrument The capacit...

Page 3: ...20nF 20mF 2000pF 2mF 200pF 200m F 200pF 20m F 200W 200MW 20W 200MW 20W 20MW 20W 2MW C R DCR D Q q Test Level Range mode Equivalent circuit 0 6 Vrms Auto and Hold Parallel and Series 4 Specifications C...

Page 4: ...ity factor ESR equivalent series resistance Rp equivalent series parallel resistance q phase angle 6 Explanation on Front Panel The front panel is shown as figure 1 1 Mini USB interface Connect with P...

Page 5: ...y When sorting mode is active press SETUP key to modify the reference value range and the tolerance setting sequentially 11 SER PAL key Used to select series and parallel mode 12 FREQ key Press FREQ k...

Page 6: ...ay is quality factor Q C D Capacitance measurement the parameter on sub display is dissipation factor D R Resistance measurement DCR DC resistance measurement mode L C R measurement readings can be po...

Page 7: ...bers the current readings on primary display called initial value when pressing the REL key and 12 b When using insert the black and red testing lines with alligator clip into the terminal and termina...

Page 8: ...gs to the component type press FUNC key to select L C or R measurement mode 2 Insert the standard component into the input therminal Press SORTING key to enter into the sorting mode and the Sorting sy...

Page 9: ...nce setting mode automatically 6 When setting the tolerance press D Q ESR key SER PAL key to select tolerance range 0 25 0 5 1 2 5 10 20 80 20 The default tolerance is 1 Press ENTER key to confirm 16...

Page 10: ...x Lx Cx LCUR LPOT HPOT HCUR GUARD GUARD INPUT Rx Lx Cx LCUR LPOT HPOT HCUR Figure 8 Figure 9 18 4 The default test mode is Auto LCR mode the capacitance value will be shown on primary display and the...

Page 11: ...ilable 3 The default test mode is Auto LCR mode the resistance value will be shown on primary display and the phase angle q will be shown on secondary display In Auto LCR mode the D Q ESR key SEL PAL...

Page 12: ...t and then firmly insert DC plug into DC input end in the right of the meter When unplugged firstly pull out the DC plug perpendicular to DC input end and then unplug the AC plug from the electrical o...

Page 13: ...s warranty on new and unused products to end user customers only but have no authority to extend a greater or different warranty on behalf of KUSAM MECO KUSAM MECO s warranty obligation is limited at...

Reviews: