*As Hazardous-Alert: When making resistance measurements in
TM
AutoCheck mode, an unexpected display of voltage readings
alerts you that the object under test is being energized.
*Ghost-voltage Buster: Ghost-voltages are unwanted stray
signals coupled from adjacent hard signals, which confuse
TM
common multimeter voltage measurements. Our AutoCheck
mode provides low (ramp-up) input impedance (approx. 3k at
W
low voltage) to drain ghost voltages leaving mainly hard signal
values on meter readings. It is an invaluable feature for precise
indication of hard signals, such as distinguishing between hot and
open wires (to ground) in electrical installation applications.
TM
AutoCheck mode input impedance increases abruptly from
initial 3k to a few hundred k 's on high voltage hard signals.
W
W
"LoZ" displays on the LCD to remind the users of being in such low
impedance mode. Peak initial load current, while probing
1000VAC for example, can be up to 471mA (1000V x 1.414 / 3k ),
W
decreasing abruptly to approx. 3.1mA (1000V x 1.414 / 460k )
W
TM
within a fraction of a second. Do not use AutoCheck mode on
circuits that could be damaged by such low input impedance.
Instead, use rotary selector V or V high input impedance voltage
modes to minimize loading for such circuits.
WARNING:
19
20
+ACV
+ACV
DC + ACV
, DCV, DCV
functions
Press the
SELECT
button momentarily to select the subject
functions in sequence. Last selection will be saved as power up
default for repeat measurement convenience.