21
+Hz
+ACmV
ACmV , Hz
functions
Press the SELECT
button momentarily to select the subject
functions in sequence. Last selection will be saved as power up
default for repeat measurement convenience.
+ACmV
+ACmV
DCmV, DCmV
, DC+ACmV
, Logic-Level Hz
& Duty%
Press the
SELECT
button momentarily to select the subject
functions in sequence. Last selection will be saved as power up
default for repeat measurement convenience.
22
CAUTION
Using resistance and continuity function in a live circuit will
produce false results & may damage the instrument. In many
cases the suspected component must be disconnected from the
circuit to obtain an accurate reading.
nS Conductance,
W
Resistance, Continuity functions
Press the
SELECT
button momentarily to select the subject
functions in sequence. Last selection will be saved as power up
default for repeat measurement convenience.
Conductance (unit-Siemens) is the inverse of Resistance, that is
S=1/
W
or nS=1/G
W
. It virtually extends the Resistance measure-
ments to the order of Giga-Ohms for leakage measurements.
Note :
Continuity function is convenient for checking wiring
connections and operation of switches. A continuous beep tone
indicates a complete wire.