6.3.6 Replacing the handle
An adhesive (e.g. hair spray) is used in these instructions. When applied to the handle, this substance works as a
lubricant and as an adhesive once dry.
CAUTION!
Risk of accidents due to the handle comes loose
– After drying, the adhesive used must be able to resist a pull-off force of 750 N. If in doubt, contact Invacare.
■■□
1.
Remove the old handle.
2.
Remove any residue (residual adhesive, grease, dust) from the push
handle tube.
3.
Apply a thin layer of hair spray all over the surface of the push
handle tube onto which the handle is to be slid.
4.
Apply a thin layer of hair spray to the inside of the handle.
5.
Slide the new handle onto the push handle tube.
6.
Move the handle into the correct position (grooves facing upwards).
If a long handle has been fitted and this is to be replaced with a short
one, the push handle tube must be shortened by 35 mm.
The push handle tube must be replaced when switching from a short to
a long handle.
6.3.7 Replacing the foldable handle
Hole punch pliers (6mm) / Allen key (3 mm, 4 mm)
■□□
1.
Remove the old foldable push handle.
2.
Pull down the backrest cover
F
on the telescopic tube, until its
hole
B
is uncovered.
IMPORTANT!
– Make sure that the threaded insert
E
(part no. 1580450) supplied
with the new push handle is used for assembly.
3.
Place the threaded insert
E
in the telescopic tube.
4.
Punch a hole through the backrest cover with a distance of 10 mm
from the upper edge, using hole punch pliers (see graphic below).
5.
Slide the new foldable push handle
A
onto the telescopic tube.
6.
Pull up the backrest cover, until it covers completely the rear hole
in the push handle.
7.
Install the foldable push handle with screw
C
and washer
G
.
8.
Check screws
D
on both sides of the push handle and re-tighten if
necessary.
9.
Carry out the same steps for the other push handle.
IMPORTANT!
– Make sure that the folding force is approximately 5 N (0.5 kg).
IMPORTANT!
– Fixing screw
C
may only be used once. Alternatively the screw can be
cleaned (remove old thread locking adhesive) and reinstalled with new
low-strength thread locking adhesive.
The retrofit of foldable push handles requires new tubing.
10 mm
22
1659339-A