If abnormal conditions of any sort are noted (such as a faulty display,
unexpected readings, broken case, cracked test leads, etc) do not use the
tester and return it to your distributor for repair.
9
.
For safety reasons only use accessories (test leads, probes, fuses, cases,
etc) designed to be used with this instrument and recommended by
KYORITSU. The use of other accessories is prohibited as they are unlikely
to have the correct safety features.
10
.
When testing, always be sure to keep your fingers behind the safety
barriers on the test leads.
11
.
During testing it is possible that there may be a momentary degradation of
the reading due to the presence of excessive transients or discharges on
the electrical system under test. Should this be observed, the test must be
repeated to obtain a correct reading. If in doubt, contact your distributor.
12
.
The sliding shutter on the back of the instrument is a safety device. The
instrument should not be used if it is damaged or impaired in any way, but
returned to your distributor for attention.
13
.
Do not operate the function selector whilst the instrument is connected to
a circuit. If, for example, the instrument has just completed a continuity
test and an insulation test is to follow, disconnect the test leads from the
circuit before moving the selector switch.
14
.
Do not rotate function dial when test button is depressed. If the function
switch is inadvertently moved to a new function when the test button is
depressed or in lock-down position the test in progress will be halted. To
reset, release test button and press again to restart testing on new
function.
15
.
Use a damp cloth and detergent for cleaning the instrument. Do not use
abrasives or solvents.
16
.
Before using the instrument, please insert the batteries rightly in
accordance with the "9. BATTERY REPLACEMENT" of page 24.
When not installing batteries, all the functions of the instrument don't
work.