6
—
Open Circuit
Voltage(DC)
Short Circuit
Current
Range
Accuracy
Greater than 4V
Greater than
200mA
20/200Ω
Auto-Ranging
Up to 2Ω
±
(3%rdg+4dgt)
Over 2Ω
±
(3%rdg+3dgt)
Operating range(as in IEC61557) 20Ωrange
:
0.2Ω
〜
19.99Ω
/
200ΩRange
:
20Ω
〜
199.9Ω
Insulation Resistance (to IEC61557-2)
Open Circuit
Voltage (DC) Rated Current
Range
Accuracy
500V
+
20%−0%
1mA or greater
@ 500kΩ
20/200MΩ
Auto-Ranging
±
(3% rdg +3dgt)
Operating range(as in IEC61557) 20MΩ Range
:
0.5MΩ
〜
19.99MΩ
/
200MΩ Range
:
20MΩ
〜
100MΩ
Loop Impedance (to IEC61557-3)
Rated Voltage
(AC)
Nominal Test
Current at 0Ω
External Loop
Range
Accuracy
230V
+
10%
-
15%
50Hz
25A/10ms
20Ω
±
(3%rdg +8dgt)
15mA/350ms max.
2000Ω
±
(3%rdg +8dgt)
Operating range(as in IEC61557) 20Ω Range
:
2Ω
〜
19.99Ω
/
2000Ω Range
:
100Ω
〜
1999Ω
RCD testing (to IEC61557-6)
Function
Rated Voltage
(AC)
Trip Current
Settings
Trip Current
Duration
Accuracy
RCD
×
1/2
230V
+
10%
-
15%
50Hz
10/30/100/300/
500mA
2000ms
Trip Current:
-
8%
〜-
2%
Trip Time
±
(1%rdg+3dgt)
RCD
×
1
Trip Current:
+
2%
〜+
8%
FAST
150mA
50ms
Voltage Measurement
Rated Voltage
Measuring Range
Accuracy
100
〜
250V 50Hz
100
〜
250V 50Hz
3%rdg
3. SPECIFICATION
Measurement Specification
Continuity (to IEC61557-4)