INSTRUCTION MANUAL

KEW MATE 2000A/2001A

DIGITAL MULTIMETER 

WITH 

AC/DC CLAMP SENSOR

KYORITSU ELECTRICAL INSTRUMENTS

WORKS, LTD.

This 

Safety 

contains 

safe 

● Read 

● 

● Be  sure 

● Be sure t

Failure  to 

damage  to 

Failure  to 

The symbol 

parts of the 

  

  

  

Following 

This 

● Never 

● Do  not 

● 

● 

● Never 

● The 

   

WARNING

● Never  attempt  to  make  any  measurement,  if  any  abnormal  conditions  are  noted, 

such as broken case, cracked test leads and exposed metal parts.   

● Do not turn the Function Selector Switch while the test leads are connected to the 

circuit under test. 

● Do not install substitute parts or make any modification to the instrument.  Return 

the instrument to Kyoritsu or your distributor for repair or re-calibration.

● Do not try to replace the batteries if the surface of the instrument is wet. 

● Always disconnect the clamp sensor and the test leads from the circuit under test 

and  switch  off  the  instrument  before  opening  the  battery  compartment  cover  for 

battery replacement.

● Stop using the test lead if the outer jacket is damaged and the inner metal or color 

jacket is exposed.

   

CAUTION

● Make sure that the Function Selector Switch is set to an appropriate position before 

making measurement.

● Always  make  sure  to  place  the  test  leads  in  the  test  lead  holder  before  making 

current measurement. 

●Do not expose the instrument to the direct sun, extreme temperatures or dew fall.  

● Be sure to set the Function Selector Switch to the "OFF" position after use. When 

the  instrument  will  not  be  used  for  a  long  period  of  time,  place  it  in  storage  after 

removing the batteries.

● Use a damp cloth and detergent for cleaning the instrument.  Do not use abrasives 

or solvents.

●Keep your fingers and hands behind the protective fingerguard during measurement.

Measurement Category:

    To  ensure  safe  operation  of  measuring  instruments,  IEC  61010  establishes  safety 

standards for various electrical environments, categorized as O to CAT IV, and called 

measurement  categories.  Higher-numbered  categories  correspond  to  electrical 

environments  with  greater  momentary  energy,  so  a  measuring  instrument  designed 

for CAT III environments can endure greater momentary energy than one designed 

for CAT II.

   O 

:  Circuits which are not directly connected to the mains power supply. 

   CAT II  :  Electrical circuits of equipment connected to an AC electrical outlet by a 

power cord.

   CAT III  :  Primary  electrical  circuits  of  the  equipment  connected  directly  to  the 

distribution panel, and feeders from the distribution panel to outlets.

   CAT IV  :  The circuit from the service drop to the service entrance, and to the power 

meter and primary over-current protection device (distribution panel).

2.FEATURES

● Permits  AC/DC  current  measurement  up  to  60A  using  a  clamp  sensor  that  comes  

standard with the instrument

●Clamp sensor for ease of use in crowded cable areas and other tight places

● Permits current measurement with an open current-clamp sensor that does not require 

opening and closing operations by the user

●Auto-power-save function

●Buzzer for easy continuity checking

●Data hold function to freeze the readings

●LCD with a 3400 count full scale bar graph

●Shock absorbing holster for ease of storage

● Designed  to  international  safety  standard  IEC61010-1:  measurement  category 

CAT 

, 300V and pollution degree 2.

O: Device which is 
     not directly 
     connected to the 
     mains power supply

●Location for use 

 Indoor use, Altitude up to 2000m

●Accuracy-insured 

23℃±5℃, relative humidity 75% or less 

 Temperature and 

(without condensation)

 Humidity Ranges

●Operating Temperature 

0-40℃, relative humidity 85% or less 

 and Humidity Range  

(without condensation)   

●Storage Temperature 

-20-60℃, relative humidity 85% or less

 and Humidity Range  

 (without condensation)

●Source 

 

Two 1.5VDC R03 (UM-4) batteries

●Current Consumption 

Approx. 10mA

●Power-save Function 

 Shifts to the power-save state about 10 minutes after the 

last switch operation

   

 

 (current consumption: approx. 10μA)   

●Overload Protection 

AC/DC current ranges:  MODEL2000A  AC/DC  72A  for 

10 seconds

            

 AC/DC current ranges:  MODEL2001A  AC/DC 

120

A  for 

10 seconds

            

 AC/DC voltage ranges: AC/DC 720V for 10 seconds

   

 

Resistance ranges: AC/DC 720V for 10 seconds

   

 

Frequency ranges: AC/DC 720V for 10 seconds

●Withstand Voltage 

 AC3470V  for  5  sec.  between  electrical  circuit  and 

housing case

●Insulation Resistance 

 10MΩ or greater at 1000V between electrical circuit and 

housing case

●Conductor Size 

MODEL2000A Approx.  6mm diameter max.

            

 MODEL2001A Approx. 10mm diameter max. 

●Dimensions   

MODEL2000A 128(L)×87(W)×24(D)mm

            

 MODEL2001A 128(L)×92(W)×27(D)mm 

●Weight 

 

MODEL2000A Approx. 210g

            

 MODEL2001A Approx. 220g

●Accessories  

Two R03 (UM-4) batteries

   

 

Instruction Manual

4.INSTRUMENT LAYOUT

Data Hold Button

MODEL2001 MODEL2000

Range Hold Button

Digital Multimeter

Function Selector Switch

Holster

Test Lead

LCD Display

AC/DC Clamp Sensor

Barrier

Protective fingerguard

DCA

Zero Adjust

Knob

5.PREPARATIONS FOR MEASUREMENT

(1)Checking battery voltage

Set the Function Selector Switch to any position other than the OFF position. 

 If the marks on the display is clearly legible without symbol 

"BATT"

 showing, battery 

voltage  is  OK.  If  the  display  blanks  or 

"BATT"

  is  indicated,  replace  the  batteries 

according to section 

8: Battery Replacement

.

NOTE

When the instrument is left powered on, the auto-power-save function automatically 

shut the power off; The display blanks even if the Function Selector Switch is set to 

a  position  other  than  the  OFF  position  in  this  state.    To  power  on  the  instrument, 

turn the Function Selector Switch or press the Data Hold Button.  If the display still 

blanks, the batteries are exhausted.  Replace the batteries.

(2)Make sure that the Function Selector Switch is set to the appropriate range.

 

Also  make  sure  that  data  hold  function  is  not  enabled.    If  inappropriate  range  is 

selected, desired measurement cannot be made.

(3)Install Test Lead to the Holster on the side of body

 

It is possible to measure with seeing the LCD Display keep 

Test Lead installing to the Holster.

AC/DC

Battery Voltage Warning

Auto Range

Data Hold

Unit

Bar Graph

Buzzer

Manual

Range

Wind the lead around the folder.

Test Leads

AC/DC Clamp Sensor

● In order 

● Do not 

● When 

● 

M

(2) Turn 

(3)  

  (  When 

occurs.)

Note:  When 

(2) 

   ( 

Note:  Unlike 

● In order 

● 

  (

(2) 

black 

display.

Correct

 6-2-2 AC Voltage Measurement

(1)Set the Function Selector Switch to "  V."

  ("AC" and "AUTO" marks are shown on the LCD.)

(2) Connect the test leads to the circuit under test.

  Measured voltage value is shown on the display.

6-3 Resistance Measurement                                                

   

DANGER

● Never make measurement on circuits that are live. 

● Never make measurement with the battery compartment cover removed. 

●Keep your fingers and hands behind the protective fingerguard during measurement.

(1)Set the Function Selector Switch to "Ω/  ."

(2) Check that the display shows over-range. Short the test leads and check that the 

buzzer beeps and the display reads zero.

(3) Connect  the  test  leads  to  the  circuit  under  test.  Measured  resistance  value  is 

shown on the display.  When the measured value is below about 30Ω, the buzzer 

beeps.

Note:  When  the  test  leads  are  shorted,  the  display  may  read  a  small  resistance  value.  

This is the resistance of the test leads.  

             If there is an open in either of the test leads, "OL" is shown on the display. 

       On the 340Ω range, "  " is shown on the left side of the LCD.

6-4 Frequency Measurement

   

DANGER

●  In order to avoid possible shock hazard, never make measurement on circuits with a 

maximum  voltage  difference  of  600VAC/DC  or  greater  between  conductors 

(300VAC/DC or greater between a conductor and ground). 

●  Do not make measurement with the test leads connected to the circuit under test. 

Never make measurement with the battery compartment cover removed.

●  Do not make current measurement with the test leads connected to the circuit under 

test.

●Keep your fingers and hands behind the protective fingerguard during measurement.

(1)Set the Function Selector Switch to "Hz."

(2) Measuring frequency of current:

   Adjust one of the conductors to the center of the clamp sensor's arrow. Measured 

value is shown on the display.

  Measuring frequency of voltage:

   Connect the test leads to the circuit under test.  Measured frequency is shown on 

the display.

Note:  Measuring range of current frequency is 0-10kHz with minimum measurable input 

of MODEL2000A 15A(Typ)/MODEL2001A 25A(Typ). 

       Measuring range of voltage frequency is 0-300kHz with minimum measurable input 

of 30V(Typ).

     When measuring frequency, do not attach the clamp sensor and the test leads to 

the circuit under test simultaneously.

Prohibition

 

 

 

 

 

 

 

 

  

 

 

 

 

92-2318A

6-18

DISTRIBUTOR

Kyoritsu  reserves  the  rights  to  change  specifications  or 
designs  described  in  this  manual  without  notice  and  without 
obligations.

Test  Lead  Cap:Test  leads  can  be  used  under  the  CAT.II  and  III  environments  by 

attaching  a  Protective  cap  as  illustrated  below.  Use  of  our  Protective  cap  offers 

different lengths suitable for the test environments.

Barrer

Protective fingerguard: It  is  a  part  providing  protection  against  electrical  shock  and 

ensuring the minimum required air and creepage distances.

Protective Cap

Exposed metal part

Uncapped condition for CAT.II environment

Capped condition for CAT.III environments

This  instrument  satisfies  the  marking  requirement  defined  in  the  WEEE 

Directive.  This  symbol  indicates  separate  collection  for  electrical  and 

electronic equipment.
This marking means they shall be sorted out and collected as ordained in 

DIRECTIVE.

This  directive  is  valid  only  in  the  EU.  When  you  remove  batteries  from 

this product and dispose them, discard them in accordance with domestic 

law concerning disposal. Take a right action on waste batteries, because 

the collection system in the EU on waste batteries are regulated.