KEW6305
8.10
●
To start measurement automatically at the preset time and date
1
Set the start time and date on
SET UP
range.
2
Then set the Function switch to
DEMAND
range, and press the
START/STOP
key.
3
File number screen is displayed for about 1 sec., followed by the measurement screen. The
instrument goes into stand-by mode for measurement. LED status indicator while the instrument is
in stand-by mode.
File number screen
(displayed for about 1 sec.)
Measurement screen
4
Measurement starts at the preset time and date, and the LED status indicator stop flickering and
remain on for the duration of the survey.
NOTE
* The start time and date should be set after the current time in such a way to give enough time to the
user to complete all settings before the survey starts.
* When the start time and date are set before the current time, measurement starts immediately upon
pressing the
START/STOP
key.
* When the preset start time and date come after the preset stop time and date, measurement stops right
after the start of measurement.
* Even if the start and stop time have been preset and the instrument is in standby mode, pressing the
START/STOP
key at least 2 sec. will release
the stand-by mode and start a survey in Manual mode. This renders the start/ stop time and date
settings ineffective.
LED status indicator
flickers
LED status indicator
is on.