KEW6305
4.16
“
Setting 14” Stop of continuous measurement
1 Use the
Cursor
key on the selection screen, and select “Setting 14”.
2 Press the
ENTER
key to get the instrument into setting change mode. On the LCD, measurement
start time (Setting 13) + 1 hour, is displayed and second will blink.
3 Change the time and date with
Cursor
Key.
4 Then press the
ENTER
key after making necessary change.
Example:
When start / stop time and date have been set as follows,
Setting 13 (start) = 12.08.01, 08:00:00
Setting 14 (stop) = 12.08.07, 18:00:00
the instrument automatically performs measurement during the following period.
From 8:00 on August 1, 2012 to 18:00 on August 7, 2012
NOTE:
* The stop time and date (Setting 14) should be set after the start time (Setting 13) in such a way to give enough
time to the user to complete all settings before the measurement starts.
Otherwise, an error message will be displayed on the LCD and the instrument cannot start measurement and
data recording.
When an error message appears, press the
ENTER
key and rotate the Function switch to the SETUP range to
redo settings.