KEW6305
6.12
Each time the
SAVE
key is pressed; the measured data is saved in the same file. To save the data into
another file (only when SD card is used), press the
SAVE
key again on
W
range. Then repeat the
saving procedure.
NOTE
* When the Function switch is set to OFF position before closing a file, the file remains open and is not
saved. Be sure to set it to any position other than OFF and
W
, thus closing the file.
* If the
SAVE
key is pressed continuously (2 times or more in 1sec.), the measured data may not be saved
correctly.
* The file number becomes “001” when;
(1) the file number has exceeded 999
(2) after system reset
* If the same file no. exists, the old one will be overwritten.
6.4.2 Limitations of saving
Data cannot be saved by pressing the
SAVE
key during a measurement when using:
<SD card>
* when number of files opened has exceeded 511.
* when the SD card memory capacity has been exceeded
appears and further data cannot be saved. To save further data, previously saved files should be
deleted via PC or by deleting all the data in the SD card by using “
Setting 19
”.(refer to section 4 of this
manual)
<Internal memory>
* when number of files opened has exceeded 4.
* when the capacity of internal memory has been exceeded.
appears and further data cannot be saved. To save further data, previously saved files should be
deleted by using “
Setting 21
” (refer to section 4 of this manual).