7.5
KEW6305
●
Automatic measurement
at a preset time and date
1
Preset the start time and date on
SET UP
range.
2
Set the Function switch to
Wh
range, and press the
START/STOP
key.
3
The file number screen is displayed for about 1 sec. (a file is opened), followed by the measurement
screen. The instrument goes into stand-by mode. LED status indicator flicker while the instrument is
in
stand-by
mode.
File number screen
(displayed for about 1 sec.)
Measurement screen
4
The survey starts at the preset time and date, and the LED status indicator stop flickering and are
permanently on.
NOTE:
* The start time and date should be set after the current time in such a way to give enough time to the
user to complete all settings before the survey starts.
* When the start time and date are set before the current time, measurement starts immediately upon
pressing the
START/STOP
key.
* If the preset start time and date come after the preset stop time and date, the survey cannot be done.
* Even if the start and stop time have been preset and the instrument is in standby mode, pressing the
START/STOP
key at least 2 sec. releases the stand-by mode and start a survey in Manual mode. This
renders the start/ stop time and date settings ineffective.
LED status indicator
flickers
LED status indicator
is on.