9
-
●
Current consumption (representative values at 12V of supply voltage)
Range
250V
500V
1000V
2500V
5000V
VOLT
Output at
short-circuit
For 10 sec after a test is started : 700mA,
after that : 180mA
110mA
When rated
measuring current
is outputted
380mA
/0.25M
Ω
440mA
/0.5M
Ω
510mA
/1M
Ω
670mA
/2.5M
Ω
860mA
/5M
Ω
Output at
open circuit
60mA
60mA
70mA
90mA
140mA
On stand-by
30mA
When backlight
is on
Increased by 30mA
Note) Current values in above table are all approximate values.
●
Operating time: approx. 10 hours continuous
- under a load of 100M
Ω
at 5000V Insulation resistance range
●
Accessories:
Line probe: MODEL7165A
(Including MODEL8255 Straight type prod with molded parts)
Earth cord: MODEL7224A
Guard cord: MODEL7225A
Hard Case MODEL9171
Pickel Type Prod: MODEL8019
Straight Type Prod: MODEL8254
Instruction manual
Power Adapter
●
Optional accessories: USB Communication set:MODEL8258
(USB adaptor(MODEL8212USB) with KEW Windows for
3127)
Adaptor for recorder:MODEL8302
Line probe with alligator clip:MODEL7168A
Long Line probe with alligator clip (15m): MODEL7253
Summary of Contents for KEW 3127
Page 1: ...INSTRUCTION MANUAL HIGH VOLTAGE INSULATION TESTER KEW 3127 ...
Page 49: ... 46 ...
Page 57: ... 54 MEMO ...
Page 58: ... 55 MEMO ...
Page 59: ......