-
22
-
(2) Setting procedures
Follow the procedures below
[Stand-by state]
ENTER button(
)
Setting of Measurement duration (with UP/DOWN button (
))
ENTER button (
)
Setting of Output voltage*1 (with UP/DOWN button (
))
ENTER button (
)
Setting of Breakdown/Burn (with UP/DOWN button (
))
ENTER button (
)
Setting is finished.
(*1): 500V/2500V/5000V Range only
6-5-1 PI – Polarization index
This is to check a temporal increase of leakage currents flowing on insulations
and to confirm leakage currents aren
t increased as time passes.
PI is usually determined by the insulation resistances measured 1 min and 10
min after a measurement is started. PI is dependent on the shape of insulations
and influenced by moisture absorption, therefore, a check of PI is important to
diagnosis the insulation of cables.
PI =
Insulation resistance
(10 min after a start of test)
Insulation resistance
(1 min after a start of test)
PI
4 or more
4 - 2
2.0 - 1.0
1.0 or less
Criteria
Best
Good
Warning
Bad
Summary of Contents for KEW 3127
Page 1: ...INSTRUCTION MANUAL HIGH VOLTAGE INSULATION TESTER KEW 3127 ...
Page 49: ... 46 ...
Page 57: ... 54 MEMO ...
Page 58: ... 55 MEMO ...
Page 59: ......