23
8-5 Series Interference Voltage (Earth Voltage) Measurement
Measurements are automatically started during a measurement of earth
resistances and earth resistivities, and the results can be checked on the
Result display screen.
Warning message Voltage High! is displayed on the Main display when
the earth voltage (Ust) is high.
2Ω/20Ω Range
: Voltage High! warning appears when the
voltage is 12V or higher. Earth resistance
measurement cannot be made when the voltage
exceeds 15V.
200Ω〜 200kΩ Range : Voltage High! warning appears when the
voltage is 15V or higher. Earth resistance
measurement cannot be made when the voltage
exceeds 20V.
Note) DC series interference voltages cannot be measured.
8-6 Auxiliary Earth Resistance Measurement
This instrument can measure and display the auxiliary earth resistances
(Rh, Rs). When the Rh or Rs value is more than Regulated value or 50k
Ω , a warning message
or
appear. The LCD
shows Rh=OL Ω or Rs=OL when the Rh or Rs values exceed 50kΩ
. These parameters are measured automatically at auxiliary earth resistance
measurements, and can be checked on the Result Display Screen.
Note) Rh and Rs stand for Auxiliary Earth Pole H(C) and the Auxiliary
Earth Resistance of S(P) respectively.
8-7 Connection of Earth Test Leads and Simplified Measurement Probes
Connect the Earth Test Leads and Simplified Measurement Probes to the
connectors on the instrument firmly. Otherwise, a contact failure occurs and
wrong results may be read out on the LCD.
Note) Some numbers other than OL may be displayed on the LCD when
making measurement without connecting any cord or probe at 200Ω
or upper Ranges. This is not a malfunction.