5-4. DC Voltage Measurement

DANGER

 Never make measurement on a circuit in which 

voltage  over  DC600V  exists  to  avoid  getting 

electrical shock.

 Do  not  make  measurement  with  the  Battery 

Cover removed.

 

Keep  your  fingers  behind  the  barrier  on  the 

instrument during measurement.

(1) Set the Function Switch to  "

DCV

"  position. 

(2)  Connect the red test lead to V/

Ω

 terminal and the 

black test lead to COM terminal.

(3)  Connect  the  red  and  black  test  leads  to  the 

positive  (+)and  negative  (-)  sides  of  the  circuit 

under  test  respectively.  Take  the  reading  on  the 

display.  If  the  connection  is  reversed,  the  display 

indicates the  "-"  mark.

5-5. Resistance/ Cont/ Diode Measurement

DANGER

Never use the instrument on an energized circuit.

 Do  not  make  measurement  with  the  Battery 

Cover removed.

Resistance

(1)  Set  the  Function  Switch  to    "

Ω/  Cont/  Diode

"  

position.

(2)  Connect the red test lead to V/

Ω

 terminal and the 

black  test  lead  to  COM  terminal.  Confirm  "OL"  is 
indicated on the display, and then short-circuit the 
tips of test leads to make the indication zero.

(3)  Connect  the  test  leads  to  the  both  ends  of  the 

resistor under test.

(4) Take the reading on the display.

     

CAUTION

 

Even if short the test lead tips, indicated value 
may  not  be  zero.  But  this  is  because  of  the 
resistance of test leads and not a failure. 

 When test leads are open,  "OL"  is indicated on 
the  display.

Continuity

(1)  Set  the  Function  Switch  to    "

Ω/  Cont/  Diode

"  

position.  "

Ω

"    has  been  selected  by  default;  press 

the SELECT  key to change it to  "

Continuity

.

Resistance  

  Cont  

  Diode

(2)  Connect the red test lead to V/

Ω

 terminal and the 

black  test  lead  to  COM  terminal.  Confirm  "

OL

"  is 

indicated  on  the  display  and  short-circuit  the  tips 

of  test  leads.  Indication  should  become  zero  and 
buzzer sounds.

(3)  Connect  the  test  leads  to  the  both  ends  of  the 

conductor  under  test.  The  buzzer  sounds,  if  the 
resistance under test is 100

Ω

 or less.

Diode

(1)    Set  the  Function  Switch  to  "

Ω/  Cont/  Diode

position.  "

Ω

"  has  been  selected  by  default;  press 

the SELECT key to change it to "

Diode

Resistance  

  Cont  

  Diode

(2)  Connect the red test lead to V/

Ω

 terminal and the 

black test lead to COM terminal.

(3)  Connect the red and black test leads to the Anode 

and  Cathode  of  the  diode  under  test  respectively. 
Take the reading on the display. If the connection 
is reversed, the display indicates "

OL

.

CAUTION

 Some of diodes cannot be tested. Indication on 
the display will be "OL" .

 (Zener diode, LED and 

so on)

6. Other functions

6-1. Sleep Function

(1)  This  is  a  function  to  prevent  the  instrument  from 

being left powered on in order to conserve battery 
life. 

      This  function  causes  the  instrument  to  enter 

Sleep  mode  about  15  minutes  after  the  last 
key  operation.  To  exit  the  Sleep  mode,  turn  the 
Function switch to "

OFF

", then to any other position, 

or press any Key.

(2)  Sleep Function is disabled when; MIN/MAX
      Function  is  selected.  Continuous  measurement  is 

made  with  the  Sleep  Function  being  disabled.  To 
activate  Sleep  Function  again,  disable  the  MIN/
MAX Function.

CAUTION

 The  instrument  consumes  small  amount  of 
battery  power  in  the  Sleep  mode.  Set  the 
Function Switch to the OFF position after use.

6-2. HOLD Key

(1) Data Hold Function
     This  is  a  function  to  freeze  the  measured  value 

on the display. Press the "HOLD" key to freeze the 
reading. 

     The  reading  will  be  held  regardless  of  subsequent 

variation in input. "

H

" is indicated on the upper left 

corner of the display while the instrument is in the 
Data Hold mode. To exit Data Hold mode, press the  
"HOLD"  key again.

    

CAUTION

 Held  readings  are  released  when  Sleep 
Function is activated while the instrument is in 
the Data Hold mode.

(2) Backlight ON/OFF (KEW2055 only)
     Pressing the HOLD key 2 sec or more lights up the 

Backlight.  Pressing  the  HOLD  key  2  sec  or  more 
again turns off the Backlight.

6-3. NCV Function

Red  LED  on  the  upper  area  on  the  Panel  lights  up 

at  all  functions  except  for  OFF  when  electric  field 

exceeding  100V  is  detected  by  the  sensor  installed  in 

the Jaws. 

It  indicates  a  presence  of 

voltage in an electrical circuit 

or  equipment  without  touching 

them.

N C V   S e n s o r   c a n   d e t e c t 

electrical  field  only  from  the 

direction indicated in the right 

figure.

Put  the  fixed  element  (left 

side)  closer  to  the  conductor 

under test.

Detection  against  in-wall 

outlet is impossible.

DANGER

 

The  LED  may  not  light  up  due  to  installation 

condition  of  electrical  circuit  or  equipment.   

Never  touch  the  circuit  under  test  to  avoid 

possible  danger  even  if  the  LED  for  NCV 

doesn t light up.

 

Check the functionality of LED on a well-known 

power  supply  prior  to  measurement.  When  the 

LED doesn t light up, do not make measurement.

 NCV indication is affected by external voltage, 

how to hold or place the instrument.

6-4. MIN/MAX Function 

CAUTION

 SELECT,  ZERO,  Hz/DUTY  keys  are  disabled 

whileMIN/MAX Function is being activated.

(1)  

AC/DC Current Range

(AC600A only on KEW2040)

     Pressing  the  MIN/MAX  Key  at  600A  &  1000A 

Function  enables  min  or  max  value  measurement. 

Press the MIN/MAX Key to select MAX or MIN. The 

max or min value within measuring range is being 

held  until  this  function  is  disabled.  "MIN"  or  "MAX"  

is  indicated  on  the  display  while  this  function  is 

being  activated.  To  disable  this  function,  press 

down  the  MIN/MAX  Key  at  least  2  sec  or  change 

functions.

(2) 

AC/DC Voltage Range

 

CAUTION

 Pressing  the  MIN/MAX  Key  without  applying 

voltage disables the Auto-ranging function and 

fixes  the  Range  to  6V.  Connect  the  test  leads 

to the circuit under test and press the MIN/MAX 

Key  after  an  appropriate  range  is  selected  by 

Auto-ranging function.

Pressing  the  MIN/MAX  Key  enables  min  or  max  value 

measurement.  Press  the  MIN/MAX  Key  to  select  MAX 

or MIN. The max or min value within measuring range 

is  being  held  until  this  function  is  disabled.  "MIN"  or   

"MAX" is indicated on the display while this function is 

being activated.

To disable this function, press down the MIN/MAX Key  

at least 2 sec or change functions.

6-5. ZERO Function 

CAUTION

 

MIN/MAX, PEAK keys are disabled while ZERO 

Function is being activated.

Zero Adjustment Function at Current Range "

Δ

" mark 

is to be indicated at the upper right on the display while 

ZERO function is being operated. Indication of relative 

value  at  Current,  Voltage,  Resistance:  Pressing  the 

ZERO  Key  indicates  REL  (relative  value)  Press  the 

ZERO  Key  to  save  the  initial  value  at  the  start  of 

measurement as a reference value. Then the difference 

between the later measured values and the reference 

value is indicated on the display. 

The  Auto-ranging  function  is  disabled,  while  this 

function is being activated, and the Range is fixed to 

the Range selected at the start of measurement. 

Relative value is indicated within following ranges.

Measuring range

=(Full-scale value at the fixed Range) ‒ (Initial value)
To disable this function, press down the MIN/MAX Key 

at least 2 sec or change functions.

6-6. Over-flow indication 

When  the  input  exceeds  the  measuring  range  at  each 
Function other than Voltage and 1000A Range, "OL" or 
"-OL" is indicated on the display.

7. Battery Replacement

WARNING

 

MIN/MAX, PEAK keys are disabled while ZERO 

Function is being activated.

CAUTION

 

Do not mix old and new batteries.

 

Install batteries in correct polarity as indicated 

in the Battery Compartment.

Replace  the  batteries  when  a  Low  Battery 

Voltage  warning  "BATT"  mark  is  indicated  on  the 

display.  Note  that  when  the  battery  is  completely 

exhausted, the display blanks without "

BATT

" mark 

shown.

(1) Set the Function Switch to  "OFF"  position.

(2)    Unscrew  and  remove  the  Battery  Compartment 

Cover on the bottom of the instrument.

(3)    Replace  the  batteries  observing  correct  polarity. 

Use new R03 (AAA) or LR03 / 1.5V batteries.

(4)    Install  the  Battery  Compartment  and  tighten  the 

screws.

display.

(2)  With  the  transformer  jaws  closed  and  without 

clamping  them  onto  the  conductor,  press  the  

"

ZERO

"  key  to  zero  adjust  the  display.  (mark  is 

displayed at the upper right on the display.)

(3)  Press the trigger to open the transformer jaws and 

clamp them onto the one conductor under test, the 

conductor should be at the center of the jaws, then 

take the reading on the display.

(4)  Set the Function Switch to an appropriate position 

according to current under test.

(5)  Pressing  the    "

ZERO

"    key  again  releases    "

ZERO

"  

function.  (mark  at  the  upper  right  on  the  display 

disappears.)

CAUTION

 When  the  current  flows  from  the  upside  (the 

display side) to the underside of the instrument, 

the polarity of the reading is positive and vice 

versa.

5-3. AC Voltage Measurement

DANGER

 Never make measurement on a circuit in which 

voltage  over  AC600V  exists  to  avoid  getting 

electrical shock.

 Do  not  make  measurement  with  the  Battery 

Cover removed.

 Keep  your  fingers  behind  the  barrier  on  the 

instrument during measurement.

(1)  Set the Function Switch to  "

ACV

"  position.

(2)  Connect the red test lead to V/

Ω

 terminal and the 

black test lead to COM terminal.

(3)  Connect  the  test  leads  to  the  circuit  under  test. 

Take  the  reading  on  the  display.  Pressing  the  

"

Hz/DUTY

"    key  while  reading  is  indicated  on 

the  display  switches  the  indication  in  following 

sequence.

AC Voltage  

  Hz  

  DUTY

CAUTION

Hz/DUTY Function requires AC40V or higher. 

 To  measure  a  frequency,  measure  the  voltage 

on the electrical circuit in advance.

 Then  press  the  Hz/DUTY  key  to  enter  into 

frequency measurement.

 Readings  of  frequency  may  fluctuate  or  be 

influenced under noisy environment.

 

Anode

 

Cathode

 

92-1801B

07-01

8. Maintenance

Cleaning

Use  a  cloth  dipped  in  water  or  neutral  detergent  for 

cleaning the instrument.

Do not use abrasives or solvents. Otherwise, instrument 

get damaged, deformed or discolored. 

Reviews: