11
25-) Check if there is any oversaturation. We can check the Value across the image
quickly. Any value below 65000 is good. Over 65000 means oversaturation. Therefore,
image can’t be used and taken again.
26-) If everything is okay, take 4 more exposures with 20 seconds exposure time and
check all for oversaturation as well.
27-
) Back to the big screen on the left. From Lamp Control section click “OFF” for Flat-
Lamp.
28-) Again from the same Lamp Control Section, choose Ne-
Lamp and click “ON”.