background image

IM-T-277F  Jul. 2021 

 

 

DTK-A Small-size Displacement Transducer INSTRUCTION MANUAL 

Thank you for purchasing this KYOWA product. Before using it, please read this 
instruction manual carefully. Also, keep the manual within easy reach so that you 
can refer to whenever necessary.   
Specifications  and  dimensions  described  in  this  manual  could  be  changed 
without notice. Please visit our website for the latest version. 

1. Calling the operator's attention 

The following cautionary symbols and headlines are used to invite the operator’s 
attention.  Be  sure  to  observe  the  accompanying  precautions  in  order  to 
safeguard the operator and preserve the performance of the instrument. 

Warning

 

Improper operation of the system may result in severe 
injury of the operator. 

Caution

 

Cautions are given to invite the operator’s attention, in 
order to avoid instrument failure or mal-function. 

2. Important notice   

Unless specified,  strain-gage  transducers  must not  be  used  under  hydrogen 
environment. 

3. Safety precautions 

Warning

 

 

As you push the rod inward, the rod returns to its initial position by reaction 
force. Handle the product with care to avoid eye poking. 

4. Handling precaution 

Caution

 

 

Do not apply excessive axial or bending force on the rod. 

 

When  changing  the  gauge  head  or  screws,  be  sure  to  insert  the         

accessory pin(

φ

2×20)through the hole, located on the cable side of the 

rod, to secure the rod in place. Or, the transducer will be damaged.

 

 

 

 

 

 

(Note) 
Do  not  turn  the  gauge  head  while 
holding the body. 

 

Do not disassemble the product. 

 

Do not use the product under water and dusty environment. 

 

Avoid usage in vibration. 

 

Pay similar attentions toward the product as one gives to regular dial gages. 

 

Always keep the rod clean. 

 

Make sure that the bending radius of cable is longer than 10 times of a 

diameter of the cable.

 

5. Installation 

5.1 

Fix the transducer to an immovable point as illustrated, using an M6 bolt 

and the accessory plate, two M3 bolts. So, arrange the M6 bolt by yourself. 

 
 
 
 
 
 
 
 
 
 

5.2 

Initial unbalance with the rod fully extended is approximately -2.5 to -0.5 

mV/V. 

5.3 

Make sure the displacement of about 0.3 mm is applied to the transducer. 

5.4 

Same as other dial gauges, the probe is made to contact the measuring 

point for measuring. However, when a measuring dynamic phenomenon, 
that the rod may fail to respond to it on occasion. So make sure before 
measuring that it can respond. 

5.5 

To fix the rod to the measuring point, remove the gauge head then fix to the 

measuring point, using the M2.5 screw or the M2.5 screw which is on the 
other end of the transducer. 

6. Connection 

6.1 

Connect the transducer to a measuring instrument.   

6.2 

When using a measuring instrument other than KYOWA, connector plug as 

follows. 

 
 
 
 
 
 
 
 
 

6.3 

After the power ON, always preheat the product for approximately 5 to 10 

minutes. 

7. Conversion 

7.1 

Use the calibration constant described on the Test Data Sheet to convert a 

measured value into a displacement value. 

7.2 

When a strain amplifier is in use, output reads in ×10

-6 

equivalent strain. Find 

a  displacement  value  corresponding  to  ×10

-6

  strain.  Then,  obtain  a 

displacement value through multiplication using the following equation. 

Displacement (mm) = Strain amplifier’s output (

×

10

-6

 strain) 

x Calibration constant (mm

1

×

10

-6

 strain) 

7.3 

When  using  an  amplifier  of  other type  or  a recorder,  first  find  the  exact 

bridge exciting voltage applied. Second, find the displacement value that 
corresponds to 1(μV) output voltage against 1(V) bridge excitation voltage. 
Then,  obtain  the  displacement  value  through  multiplication  using  the 
following equation. 

Displacement (mm) = 

Bridge output voltage (μV) 

×Calibration constant (mm

1μV/V) 

Bridge excitation voltage (V) 

8. Storage precautions and inspection 

8.1 

Avoid water, dust and oil from the product. 

8.2 

For storage, be sure to put the accessory protection tube on the rod. 

8.3 

Recommend  calibrate  the  product  once  a  year  or  so.  (Contact  your 

KYOWA representative.) 

8.4 

If an abnormal initial value or reading appears, measure input resistance, 

output resistance as well as insulation resistance (which should be 100M

 

or higher). 
If abnormal resistance is found, the cause may be failure of the sensing 
element. In this case, contact your KYOWA representative. 

Caution

 

 

To measure insulation resistance, apply a voltage lower than 50V to the 
insulation resistance tester. 

Pin(φ2×20) 

Washer 

Spring washer 

Rod 

Gauge head 

NG! 

Reviews: