117
The [MOD] is for the RS-232C only. You are not able to use the [MOD] with the RS-485.
When sending the [MOD] on the RS-485, the product replies the ng<CR><LF>.
(4) [MOD]
Content:
Set the transmission mode.
Format:
MODpp<CR>
Parameter:
pp
00: Continuously outputs the measured value and status.
02: Outputs the display value and status only 1 time while holding.
10: Operation with the control command.
Description and others
This command becomes disabled with the power OFF.
Power ON and it is in Mode 10 operation mode.
Return value format:
(1)Return value format of the mode 00
D1±xxxxxx<CR><LF>
Sabcdefghi<CR><LF>
・
The mode 00 repeatedly outputs the measured value and status in order.
・
The ±xxxxxx is a measured value (-99999 to 99999) with a decimal point included.
If the indicated value has no decimal point, it is filled with ‘0s’ from the head.
・
Return value abcdefghi represents status with ‘1’ or ‘0’.
Following is the status for outputting “1.”
a: HH comparator is ON.
b: HI comparator is ON.
c: OK comparator is ON.
d: LO comparator is ON.
e: LL comparator is ON.
f: Measuring
g: Detecting
h: holding
i: Errors
・
Set the mode 10 to stop the outputting.
・
When you cannot set the mode.
Turn OFF the WGA-910A once and turn ON again to set it to mode 10.
(2) Return value of the mode 02
D1±xxxxxx<CR><LF>
Sabcdefghi<CR><LF>
・
The mode 02 outputs the measured value and status once during the holding time.
・
After the data is output, the WGA-910A is automatically set to the mode 10.
・
The contents of the measured value and status are the same with the mode 00.
・
When the mode 02 is specified while the WGA-910A is moving with the mode 00,
the ng<CR><LF> is replied.
(3) Return value format of the mode 10
ok<CR><LF>
・
The WGA-910A is controlled with commands.
・
The WGA-910A is controlled with the contents of the control commands.
(5) [/M]
Content:
Setting saving command (Saves the setting value to the internal memory.)
Format:
********/M<CR>
Corresponding command
Return value format:
Add ‘/M<CR>’ at the end of the determined command. Then, it is stored in the internal
memory.
If the power is turned OFF, the contents determined without this command shall return
to the original setting.
(Only commands with asterisk mark (*).)
NOTE