122
8-3-4. Setting
(17) [SBV]
Content:
Set the excitation voltage.
Format:
SBVa<CR>
Parameter:
a
0: 2V
1: 10V
Return value format:
ok<CR><LF>
Description and others
Corresponding to the [/M] command.
Reading command
PRR012
Return value
a<CR><LF>
0: 2V
1: 10V
(18) [ZAD]
Content:
Conduct no-load zero.
Format:
ZAD<CR>
Parameter:
None
Return value format:
ok<CR><LF>
Description and others
Corresponding to the [/M] command.
(19) [VCL]
Content:
Conduct actual load calibration.
Format:
VCL ± yyyyyy, zz<CR>
Parameter:
±yyyyyy, zz
Unit: 01 to 79
79 types
Refer to “12-2 UNIT LIST.”
Rated display value: -99999 to 99999 (count)
Indicated value with a decimal point. If the indicated value has no
decimal point, it is filled with ‘0s’ from the head.
Return value format:
ok<CR><LF>
E-12<CR><LF> (When an error occurs)
Description and others
The results of the command will be saved.
Reading command
PRR010
Return value
For details, see the [EIC].
(20) [EIC]
Content:
Conduct sensitivity registering calibration.
Format:
EIC ±x.xxxx, ±yyyyy, zz<CR>
Parameter:
±x.xxxx, ±yyyyy, zz
Unit: 01 to 79
79 types
Refer to “11-2 UNIT
LIST.”
Rated display value: -99999 to 99999 (count)
Indicated value with a decimal point. If the indicated
value has no decimal point, it is filled with ‘0s’ from the
head.
Rated output value: -3.2000 to 3.2000 (mV/V)
Return value format:
ok<CR><LF>
E-12<CR><LF> (When an error occurs)
Description and others
The results of the command will be saved.
Reading command
PRR010
Return value
±x.xxxx, ±yyyyy, zz
Unit:
Rated display value
Rated output value