5. CALIBRATION
5-1. OUTLINE OF CALIBRATION
The WGA-910A has 4 calibration methods; manual input calibration, TEDS-based automatic calibration, partial
calibration based on TEDS calibration data, and numeric value registering calibration.
Manual input calibration:
Enters the calibration data manually.
The manual input calibration includes the "Sensitivity registering calibration" that
calibrates data by registering a sensitivity value written in the test data sheet of
transducers and "Actual load calibration" that calibrates data by applying the
already known load to transducers. You are able to calibrate data properly.
TEDS-based automatic calibration: Applicable to the TEDS-compatible sensors.
Calibrates the data automatically by loading all the necessary information from
the TEDS information.
Partial calibration based on TEDS calibration data:
Selects and calibrates the target calibration data only when loading the TEDS
information.
Numeric value registering calibration:
Registers the input value for indicating “0” as the initial value manually.