5-3. HOW TO CALIBRATE
5-3-1. Calibration Window
<How to operate>
1) Press the
key on the “
Measuring
” window.
2) Press the
key on the “
Setting
” window.
5-3-2. Manual Input Calibration
Conducts calibration by manual setting in order of excitation voltage, no-load zero, Sensitivity registering calibration
or in order of actual load calibration and unit.
(1) Bridge Voltage (Excitation Voltage)
Capable of setting the excitation voltage that applies to the sensor from 2 V or 10 V.
Select the higher voltage within the recommended excitation voltage of the sensor.
<Setting range>
2 V or 10 V
<How to operate>
1) Press the
key on the “
Calibration
” window.
2) Press either
or
key and press the
key to register the setting value.
3) After the registration, press the
key to move to the
“
No-load Zero
” window.
Calibration
'Bridge Voltage','No-Load Zero
,'Sens.Regist.Cal'or'Actual
Load Cal'and 'Unit'are set.
Return
Measure
Manual Input Cal.
Next
1 / 2
TEDS Auto Cal.
TEDS Part Cal.
TEDS Ope.Setting
'TEDS Read Ope.'and 'Zero dur-
ing TEDS'are set.TEDS informa-
tions are displayed.
Calibration is automatically
conducted with all reading
from TEDS built-in sensor.
Calibration is conducted with
the only selected item from
TEDS built-in sensor.