43
5-3-3. TEDS Auto Cal. (TEDS-based Automatic Calibration)
The calibration is automatically conducted by reading the required information for the calibration (applied voltage,
rated output, and unit) from the TEDS built-in sensor and the calibration data that is saved inside is overwritten.
However, the “TEDS-based Automatic Calibration” is valid when “
Allow”
is selected in the “
Key Ops.”
of the
“
TEDS Reading Operation”
window.
The
key is invalid and displayed in grey when “
Prohibit”
is selected.
In addition, the TEDS information is read after conducting the no-load zero and calibration is conducted when
selecting “
Execute”
in the “
Zero during TEDS”
window.
<How to operate>
1) Press the
key on the “
Calibration
” window.
2) After pressing the
key, the “
Confirmation of TEDS Auto
Cal.”
window appears.
3) Press the
key. After conducting the no-load zero, the TEDS
information is read and the calibration is started.
Press the
key to cancel the calibration and gets back to the
previous window.
4) The calibration starts and the “
TEDS Calibration”
window on the
upper left blinks.
5) The calibration is finished after a short time and a new calibration
value is displayed on the “
New”
column.
Press the
key at last to move to the “
Measuring
” window.
Calibration
'Bridge Voltage','No-Load Zero
,'Sens.Regist.Cal'or'Actual
Load Cal'and 'Unit'are set.
Return
Measure
Manual Input Cal.
Next
1 / 2
TEDS Auto Cal.
TEDS Part Cal.
TEDS Ope.Setting
'TEDS Read Ope.'and 'Zero dur-
ing TEDS'are set.TEDS informa-
tions are displayed.
Calibration is automatically
conducted with all reading
from TEDS built-in sensor.
Calibration is conducted with
the only selected item from
TEDS built-in sensor.
Blink