5-3-4. TEDS Part Cal. (Partial Calibration Based on TEDS Calibration Data)
Only the selected information from the TEDS built-in sensor (select one or multiples from the applied voltage, rated
display, rated output, unit) is updated the calibration data that was saved internally and calibrated automatically.
However, the “
Partial
C
alibration Based on TEDS Calibration Data”
is valid when selecting the
“Key Ope.
” in
the “
TEDS Reading Operation
” to “
Allow
.”
The
key is invalid and displayed in gray when setting “
Key Ope.
” to “
Prohibit
.”
In addition, the TEDS information is read and calibration is conducted after conducting the no-load zero when
setting the “
Zero during TEDS
” to “
Execute
.”
Select the target item for updating in the “
TEDS Cal. Item
” window.
<Setting Range>
(Common to the
Ex. Volt.
,
Rd. Out.
,
Rd. Cap.
, and
Unit
)
Execute or Prohibit
<
How to operate
>
1) Press the
key on the “
Calibration
” window.
2) Press the
key on the “
TEDS Part Cal.
” window.
3) Every time the right key of the calibration item is pressed,
or
changes.
After the setting, press the
key.
After the registration, press the
key.
4) Press the
key to display the “
Confirmation of TEDS Part
Cal.
” window.
5) Press the
key. After conducting the no-load zero, the TEDS
information is read and the calibration starts.
Press the
key to cancel the calibration and gets back to the
previous window without conducting the calibration.
Next
1 / 2
TEDS Auto Cal.
TEDS Part Cal.
TEDS Ope.Setting
'TEDS Read Ope.'and 'Zero dur-
ing TEDS'are set.TEDS informa-
tions are displayed.
Calibration is automatically
conducted with all reading
from TEDS built-in sensor.
Calibration is conducted with
the only selected item from
TEDS built-in sensor.
Calibration
'Bridge Voltage','No-Load Zero
,'Sens.Regist.Cal'or'Actual
Load Cal'and 'Unit'are set.
Return
Measure
Manual Input Cal.
MEMO