46
“
TEDS Error
” window blinks when the TEDS information cannot read correctly or TEDS information
value is out of specifications.
After stopping the reading of the TEDS information, the “
TEDS Error
” window blinks approx. 3 seconds.
After that, the current calibration value is displayed.
At this time, the calibration value is not updated.
The “
TEDS Calibration”
window blinks after the following settings or conducting the following operations
during the setting or during the setting.
<Setting>
Set “Cnt. Input”, “Power ON”, “Command” to “
Allow
” in the “
TEDS Reading Operation
” window.
<Operation>
・
TEDS command of the control input is turned ON.
・
Receives a communication command (EID) that executes the automatic sensitivity-registering calibration
with the TEDS.
・
After turning ON the power again.
6) The calibration starts and the “
TEDS Calibration”
window on the
upper left blinks.
7) The calibration finishes after a short time and a new calibration value
is displayed on the “
New”
column.
Press the
key at last to get back to the “Measuring” window.
MEMO
NOTE
Blink