(8) When displaying error.
During the TEDS calibration, due to the following causes, the TEDS information cannot be read, displays the
“
TEDS Error
” and sensitivity registering may not be conducted.
・
TEDS signal lines are wrongly connected. Check the wiring by referring to “3-2 WIRING.”
・
The connecting number of sensors is not applicable. Check that the connecting number of sensors is within
1 to 4.
・
Cable is too long.
TEDS information may not function correctly if the cable length exceeds 30 m.
Pay attention when using an extension cable to the sensor.
・
Connecting non-TEDS-compatible sensor.
After disappearing the “
TEDS Error
” display, the calibration value of the TEDS information is not applied and
the WGA-910A operates with the calibration values that are registered before the TEDS calibration.