(2) Min. Scale Zero Fix (Zero around zero)
Means least significant digit fixed at zero.
Automatically sets the display number of the least significant digit (the rightmost digit) to “0” when the measuring
value is within the setting range of + and - directions.
The Min. Scale Zero Fix does not affect to the measured value but affect to the display numbers only. It improves
the appearance of the zero point.
The additional value is displayed when it is set.
<Setting Range>
0 to 9 (count)
<How to operate>
1) Press the
key on the “
Zero Compensation
”
window.
2) Input a setting value with the numeric keyboard and press the
key to register the setting value.
6-2-3. Additional Value
Adds the setting value to the measured value and displays it.
After conducting the digital zero with the additional value set to other than “0,” the setting value is displayed.
(Displayed value) = (Measured value after conducting the digital zero) + (Additional value)
<Setting Range>
-99999 to 99999 (count)
<How to operate>
1) Press the
key on the “
Measuring
” window.
2) Press the
key on the “
Setting
” window.
3) Press the
key on the “
Input Setting
” window.
4) Input a setting value with the numeric keyboard and press the
key to register the setting value.
Zero Compensation
Additional Value
Original Value
The original value of sensor
is monitored.
A function that automatically
compensates the fine zero-
shift.
The measured value is display-
ed with adding the 'Additional
Value'.
Input Setting
A function that stables the
indicated value by reducing
unnecessary noise.
Smoothing
Return
Measure