59
The sensor output value is displayed in “--.---- mV/V” and “--.----
με” before starting the measurement.
6-2-4. Sensor Output Value (Original Value)
Displays the current output value of the sensor on the monitor.
After pressing the
key, the output value is measured for every 0.5 second and displayed in both units
“mV/V” and “με.”
<Setting Range>
-3.2000 to 3.2000 (mV/V) (-6400.0 to 6400.0 με)
<How to operate>
1) Press the
key on the “
Measuring
” window.
2) Press the
key on the “
Setting
” key.
3) Press the
key on the “
Input Setting
” key.
4) After pressing the
key, the output value of the sensor is
continuously measured and displayed for every 0.5 second.
5) Press the
key to stop the measurement of the sensor output
value.
Zero Compensation
Additional Value
Original Value
The original value of sensor
is monitored.
A function that automatically
compensates the fine zero-
shift.
The measured value is display-
ed with adding the 'Additional
Value'.
Input Setting
A function that stables the
indicated value by reducing
unnecessary noise.
Smoothing
Return
Measure
MEMO
Original Value