69
(5) Display Mode
Selects the display operation of the peak (bottom) value from the two modes.
Sets the target Display Mode. For the “Block-specified Peak", “Block-specified Bottom”, “Time-specified Peak ",
“Time-specified Bottom", “Block Averaging” and “Time Averaging" in the "Operation Mode" window only.
For other operation modes, “Display Mode" setting is invalid.
<Setting Range>
Normal Disp.
The peak (bottom) value is always updated and displayed in response to changes of the input
in the detecting block. The peak (bottom) value is held and displayed in the holding block.
Hold Disp.
The measured value immediately after turning ON the holding command is held and displayed
in the delay time and detecting block. The peak (bottom) value in the detecting block is
displayed when entering the holding block.
<How to operate>
1) Press the
key on the “
Measure Mode Setting (1/2)
”
window.
2) Press the
key on the “
Measure Mode Setting (2/2)
”
window.
3) Select and press the target key and press the
key to register
the setting value.