82
Satisfying trigger condition means starts
displaying the waveform when the
measured value passed the preset level in
the
Passed Level
and the measured
value passed as presetting in the
Level
Passed Way
.
6-4-3. Start Mode of Wave (Trigger Mode)
Sets a start mode (trigger mode) for displaying the waveform. Operations of modes are described as follows.
<Setting Range>
Auto Mode, Normal Mode, Single Mode
<How to operate>
1) Press the
key on the “
Wave Display Setting (1/2)
”
window.
2) Press the target key and
key for registration.
(1) Auto Mode
Starts displaying the waveform and automatically updates it even if the trigger conditions are not satisfied.
Press the
key or turn ON the “Waveform Command” of the control input for starting the waveform
display.
After displaying the waveform of one window to the end point of the X-axis, the waveform is held in the
“Waveform Holding Time” and then the next waveform is displayed repeatedly.
Press the
key or turn OFF
the “Waveform Command” for stop
loading.
(2) Normal Mode
Press the
key or turn ON the “Waveform Command.”
The WGA-910A becomes trigger condition waiting mode (see the above MEMO) and starts displaying the
waveform only when the measured value satisfies the trigger condition.
After displaying the waveform of
one window until the end point of
the X-axis, the WGA-910A becomes
trigger condition waiting mode and
holds the waveform without updating
it.
1 / 2
The level that measured value
passes for satisfying trigger
condition is set.
Next
1000.00kN
Auto Mode
0.00kN
The start and end point of Y-
axis are set.
The mode to start the wave
display is set.
Y-axis Setting
Start Mode of Wave
Passed Level
Wave Display Setting
The end point of X-axis(time)
is set.
10.0sec
Return
Measure
X-axis Setting
MEMO