25
3.3: Operation principle
multi-module synchronization system, the user will be able to intelligently set the control tables of
different modules, as well as the division of the reference frequency for the required input-output
processes.
!
Note that for the slave L-502 there is a frequency limitation on the input
CONV_IN - no more than 1.5 MHz. Thus, two or more L-502 slave modules
can be synchronized only at a reference frequency of 1.5 MHz from the master.
Stopping the primary synchronization scheme is done only programmatically and
asynchronously.
Узел
АЦП
Узел
ЦАП и
цифрового
вывода
Узел
цифр.
ввода
Комму-
татор
Схема
первичной
синхрони-
зации
L-502
Генератор
f
ref
f
ref
Общие условия синхронизации всех процессов ввода-вывода
в L-502 (выбор опорной частоты и условий старта)
DI_SYN1
DI_SYN2
DI_SYN1
DI_SYN2
CONV_IN
START_IN
START_OUT
CONV_OUT
К ведомому
L-502
От ведущего
L-502
Внешняя
синхронизация
DI_SYN1
DI1...DI16
DI_SYN2
Управля
ющая
таблица
DAC2
DO1...DO16
DAC1
Cхема вторичной
синхронизации
данных АЦП
Поток данных “на ввод”
Поток данных “на вывод”
Синхр.
Синхр.
Синхр.
Синхр.
X1...X16
Y1...Y16
GND32
2.0 / 1,5 МГц
Вход
Выход
I
II
Схема селекции
данных АЦП в
зависимости
от вторичных
условий
синхронизации
Fig. 3-2. L-502 synchronization system structure
3.3.5.2. Secondary synchronization.
!
The functionality of the secondary synchronization is embedded in the project,
but is not currently implemented. You can find out about the availability of this
functionality
in the sales department of L-Card
General synchronization conditions for all I/O processes
in L-502 (selection of reference frequency and start conditions)
ADC secondary
synchronization data
scheme
Node of
DAC and
digital
output
Digital
input
node
ADC
node
Switch
Generator
Control
table
Primary
synchroniz
ation
scheme
L-502
To the slave
L-502
Data flow "to output"
Data flow "to input"
Scheme of ADC
data selection in
dependence
from the
secondary
synchronization
conditions
From the
master
L-502
External
synchronization
Synch.
Input
Output
Synch.
Synch.
Synch.
MHz