Specification 

A.

 

UCP Series 

-. Dimension 

 

Name

 

UCP Ultrasonic Cleaner 

Model 

UCP-10 

UCP-20 

UCP-02 

Capacity 

10 Liter 

20 Liter 

2 Liter 

Tank Size 

(WxDxH) 

295x240x160 

500x290x160 

142x132x104 

Overall Size 

(WxDxH) 

490x368x297 

721x451x297 

290x240x200 

Temp. Range 

Amb. ~ 70 ℃ 

Time Range 

1 ~ 99 min, 4 hour 

1~99 min 

Transducer 

BLT Type 

Ultrasonic Freq. 

40 KHz 

Input Power 

AC230V, 50/60Hz or AC120V, 60Hz 

Body Material 

ABS 

Ultrasonic 

Power(Max) 

300W 

500W 

100W 

Heater Power 

190W 

320W 

Weight 

11 Kg 

17 Kg 

2 Kg 

 

-. Image of front and rear of the product (Except UCP-02) 

 
 
 
 
 
 

 
 

 

 

DOMINIQUE 

DUTSCHER SAS

Summary of Contents for UCP-02/10/20

Page 1: ...ULTRASONIC CLEANER OPERATING MANUAL ULTRASONIC CLEANER Model UCP 02 10 20 UCS 05 10 20 D O M I N I Q U E D U T S C H E R S A S...

Page 2: ...Series B UCS Series 3 Operation 8 A Function Key B Starting Method 4 Cleaning Method 10 A Direct Cleaning B Indirect Cleaning C Effective Cleaning 5 Trouble Shooting 11 6 Warranty 11 7 Membrane 12 A U...

Page 3: ...nual marks following warning indication for you to do not hurt and for property to do not damage Ignore this expression and user makes contents that danger of death or serious injury can happen in cas...

Page 4: ...the bath Otherwise it may cause the damage to the cleaner Do not operate the cleaner with wet hands Make sure of pouring solution up to the indicating level Otherwise it may cause the overheat of heat...

Page 5: ...r is being supplied Otherwise it may result in electrical shock Make sure of using the plug connected to the ground Do not clean with placing the works directly on the bath bottom Use a basket or a tr...

Page 6: ...2x104 Overall Size WxDxH 490x368x297 721x451x297 290x240x200 Temp Range Amb 70 Time Range 1 99 min 4 hour 1 99 min Transducer BLT Type Ultrasonic Freq 40 KHz Input Power AC230V 50 60Hz or AC120V 60Hz...

Page 7: ...Heater Power 150W 190W 320W Ultrasonic Freq 40 KHz Input Power AC230V 50 60Hz or AC120V 60Hz Temp Range Amb 70 Time Range 1 99 min 4 hour Body Material SUS 304 Control Microcontroller Control Method...

Page 8: ...peration of Heater is displayed as DOT LED on FND TEMP window Display whether to operate DeGAS Display whether to generate ULTRASONIC KEY SET KEY Operated as NORMAL TEMP TIME OP MODE POWER NORMAL Sele...

Page 9: ...changing into Operation Selection Mode UP DOWN Key to select the proper mode can be used OPERATION SELECTION MODE viii In OPERATION SELECTION MODE press SET KEY once more again and then goes into POW...

Page 10: ...e indication line When the beaker are put iii Pour the solution to a beaker or several one containing the works iv Put the beaker containing the works in the bath and clean it v When putting the beake...

Page 11: ...Check Time Set 2 Request A S Cleaning power is weakened 1 Solution is old pollution level is high 2 Solution is insufficient 3 Residue on bath bottom is solidified 4 The works are too much put in the...

Page 12: ...ULTRASONIC CLEANER 7 Membrane A UCS Series D O M I N I Q U E D U T S C H E R S A S...

Page 13: ...ULTRASONIC CLEANER B UCP Series D O M I N I Q U E D U T S C H E R S A S...

Page 14: ...d Kingdom Tel 44 1865 400321 E mail labcompanion medlinescientific com FAX 44 1865 400736 China Shanghai Branch A 2113 Oriental International Plaza 85 LouShanGuan Rd Changning District Shanghai China...