background image

WEEE: Disposal at end of life. 

This equipment, when being disposed of in an EU Member State, is subject to the Waste Electrical and Electronic 

Equipment Regulations (WEEE) and must be recycled and disposed of in accordance with EU directive 2002/96/EC 

as applied in local laws of that State at the time of disposal. This refrigerator contains fl ammable insulation gases 

and must be disposed of in the appropriate way by competent persons. Ensure that none of the refrigerant circuit 

pipes are damaged prior to disposal.

The symbol below on the appliance or accompanying documents indicates that this refrigerator should not be 

treated as ordinary waste but must be taken to a special collection centre for the recycling of electrical equipment. 

For more information regarding the treatment, recovery or recycling of this product, contact your competent local 

offi  ce, electrical waste disposal or vendor from whom you purchased the refrigerator.

As this item is designed for medical or scientifi c use, before presenting for collection and disposal it will be 

necessary for the user to provide documentary evidence of decontamination and/or that it is entirely safe to handle 

and dismantle outside of a controlled environment. Goods that cannot be so certifi ed fall outside the scope of the 

Directive and remain the user’s responsibility to dispose of.

This refrigerator contains a fl ammable gas. 

This refrigerator contains gas under pressure which may explode if heated.

The refrigerator MUST NOT be located in a room or area with a volume 

less than 10m

3

.

Original instructions - Version 1 12.2018

WARNINGS

Servicing and repairs must be carried out by a suitably qualifi ed engineer. 

Only engineers who have been trained in the safe handling and use of hydrocarbon 

refrigerants can work on the refrigeration system. 

If you suspect a leak contact the manufacturer or its service agents immediately.

Do not touch the refrigerant, it can cause severe eye irritation, redness, tearing, blurred 

vision, and possible freeze burns. Contact with evaporating liquid may cause frostbite or 

freezing of the skin.

If the power supply cord is damaged, it MUST be replaced by the manufacturer, its 

service agent or similarly qualifi ed persons in order to avoid a hazard.

Do not use mechanical or other means to accelerate the defrosting process. 

Keep clear of obstruction all ventilation openings in the appliance enclosure. 

This refrigerator is climate class 4 and is designed to operate at +16°C to +32°C 

ambient room temperature.

Do not damage the refrigerant circuit.

Maximum shelf loading 40 Kg/m 

2

The noise level in normal operation is lower than 70 dB.

Do not use electrical appliances inside the refrigerator.

Do not store explosive substances such as aerosol cans with fl ammable 

propellant in this appliance.

Labcold RLDF/G refrigerators are designed for the storing of pharmaceuticals, 

medicines and laboratory products and are for professional use only by suitably 

trained, competent persons.

Summary of Contents for IntelliCold RLDF 19 Series

Page 1: ...www labcold com IntelliCold Pharmacy Refrigerator Operating Instructions RLDF G 19...

Page 2: ...gas This refrigerator contains gas under pressure which may explode if heated The refrigerator MUST NOT be located in a room or area with a volume less than 10m3 Original instructions Version 1 12 20...

Page 3: ...Temperatures 12 Alarm Guide 13 Light Icon Button 14 Door Lock 14 Digital Door Lock 15 Micro SD Card 18 Caring for your Fridge 20 Troubleshooting 21 Useful Information 22 IMPORTANT These operating inst...

Page 4: ...e bottom of your fridge For models RLDF G0519 1019 and 1519 insert shelves by placing clips in the racking and resting the shelf on top When positioning the shelves please remember to keep adequate ro...

Page 5: ...like a mobile phone or tablet When you place your hand over the controller it will reveal the various icons buttons allowing you to access different functions Like a mobile phone screen it may not wor...

Page 6: ...ample d98 Adjust the digits using the arrow icons buttons as above so if it is 21st June adjust the digits to 21 When adjusted touch the tick icon button as before H L P H L P After setting the day yo...

Page 7: ...lready in the menu touch the up and down arrows as required until you get to a screen with SFt on it Touch the tick icon button as before and it will display a number Again touch the mute icon button...

Page 8: ...C where it is placed The temperature detected by this probe will display in normal operation after the t1 screen You can tell which one is which because a small to1 or to2 will appear above the temper...

Page 9: ...tten We recommend that once the date time is set it should not be altered unless absolutely necessary for example if the controller experiences a power issue and batteries aren t fitted If you are goi...

Page 10: ...ld always be investigated The alarm will also sound when this screen is displayed 1l 2h 2l l o b t1L This will display when probe t1 detects that the air temperature is below 2oC Because low temperatu...

Page 11: ...S YOU HAVE PURCHASED A MASTER KEY Opening the Door Once the refrigerator has power you can now open the door using the factory set user code of 2244 Put in the code and after every correct digit a gre...

Page 12: ...rms and fridge function and allows you to download this information as a CSV file The controller will store this information for 18 months but we recommend that you download the data once a week espec...

Page 13: ...product The contents are not packed too closely together The contents do no touch the door wall or the bottom of the fridge and are stored away from the fan As with all refrigerators your Labcold phar...

Page 14: ...or you should contact your retailer in the first instance If you bought your fridge directly from Labcold you should contact our service team on 01256 705 580 or Email them at service labcold com When...

Page 15: ...www labcold com...