background image

9

The internal light comes on when the door opens and go off when it is shut. It can  be changed 

so the light is either on or off all the time. To change the setting, access the menu in the usual 

way by holding both the arrow icons/buttons and scroll through the menu until you get to the 

Lt

 screen or hold the light icon/button which will take you straight to the 

Aut

 screen.

Touch the tick icon/button as before to access this function and the screen will show 

Aut

. This is the 

default setting described above. To change the setting to the light being on all the time touch an 

arrow icon/button until the screen displays 

on

. Once selected, hold the tick icon/button until the 

display reverts to 

Lt

To change the light operation to not coming on at all, select 

OFF 

instead and 

again hold the tick icon/button until the screen reverts to 

Lt. 

You can either touch the mute icon/

button or just leave the controller alone and the screen will return to the normal temperature display.

10

If you do not fit batteries you will need to turn this alarm to 

OFF

. To do so, hold the arrow 

icons/buttons to enter the menu. Scroll through until you get to the screen that shows 

bAt

 

and touch the tick icon/button. Touch the arrow icon/button to change the screen from 

on

 

to 

OFF 

and hold the tick icon/button until the display reverts to 

bAt

 to confirm the change. 

The battery symbol should then disappear from the temperature display screen and the alarm 

won't sound.

There is a secondary temperature probe at the back of your refrigerator. It is a ‘wandering 

probe’ you can move anywhere in the fridge, for example placing it in a simulated load. Like 

probe 

to1

 the controller will alarm if probe 

to2 

detects that the temperature is lower than 2

o

or higher  than 8oC where it is placed. The temperature detected by this probe will display in 

normal operation after the 

t1

 screen. You can tell which one is which because a small 

to1

 or 

to2

 

will appear above the temperature displayed.

Likewise you will be able to tell which probe is alarming as the high or low alarm will be 

prefixed by  

t1

 or 

t2

   Internal Light

H/L

P

H/L

P

H/L

P

H/L

P

   Battery Alarm

The battery alarm is enabled by default. This means if you put 4AA batteries (not supplied) 

in your refrigerator the alarm will not sound unless the batteries are faulty or removed.  For 

extra security, the controller will alarm if the batteries are low or removed. 

Labcold

 strongly 

recommends that you fit batteries to your 

I

ntelli

C

old® pharmacy refrigerator for maximum 

cold chain safety, however, should you decide not to you will need to disable the alarm, please 

see 

Turning the Battery Alarm Off

 opposite

The alarm will show on the normal display as a battery symbol which will alternate with a 

screen showing 

n-b

. The alarm will sound when 

n-b is 

displayed. This alarm can be muted by 

touching the mute icon/button while the 

n-b 

screen is displayed. The visual alarm will remain 

and the screen 

n-b 

will still alternate with the temperature display. Unless you install batteries 

the alarm will sound again after the time out interval is exceeded. (see page 8). 

°

°t

1

C

H/L

P

H/L

P

Battery Alarm

Turning the Battery Alarm Off

H/L

P

H/L

P

H/L

P

Temperature Probe t02

H/L

P

°

°t

1

C

°

°t

C

2

Should you wish to disable it, enter the menu by holding the two arrow icons/buttons 

simultaneously and scroll until you get to a screen that shows 

t2

.

Touch the tick icon/button to enter the 

t2

 menu and the screen will display 

on.

 Touch the 

arrow icon/button to reveal the 

OFF 

screen and hold the tick icon/button to confirm. The 

display will then return to 

t2 

and then the normal temperature display.

Temperature detected 

by probe 

to1

Temperature detected 

by probe 

to2

Summary of Contents for IntelliCold RLDF 19 Series

Page 1: ...www labcold com IntelliCold Pharmacy Refrigerator Operating Instructions RLDF G 19...

Page 2: ...gas This refrigerator contains gas under pressure which may explode if heated The refrigerator MUST NOT be located in a room or area with a volume less than 10m3 Original instructions Version 1 12 20...

Page 3: ...Temperatures 12 Alarm Guide 13 Light Icon Button 14 Door Lock 14 Digital Door Lock 15 Micro SD Card 18 Caring for your Fridge 20 Troubleshooting 21 Useful Information 22 IMPORTANT These operating inst...

Page 4: ...e bottom of your fridge For models RLDF G0519 1019 and 1519 insert shelves by placing clips in the racking and resting the shelf on top When positioning the shelves please remember to keep adequate ro...

Page 5: ...like a mobile phone or tablet When you place your hand over the controller it will reveal the various icons buttons allowing you to access different functions Like a mobile phone screen it may not wor...

Page 6: ...ample d98 Adjust the digits using the arrow icons buttons as above so if it is 21st June adjust the digits to 21 When adjusted touch the tick icon button as before H L P H L P After setting the day yo...

Page 7: ...lready in the menu touch the up and down arrows as required until you get to a screen with SFt on it Touch the tick icon button as before and it will display a number Again touch the mute icon button...

Page 8: ...C where it is placed The temperature detected by this probe will display in normal operation after the t1 screen You can tell which one is which because a small to1 or to2 will appear above the temper...

Page 9: ...tten We recommend that once the date time is set it should not be altered unless absolutely necessary for example if the controller experiences a power issue and batteries aren t fitted If you are goi...

Page 10: ...ld always be investigated The alarm will also sound when this screen is displayed 1l 2h 2l l o b t1L This will display when probe t1 detects that the air temperature is below 2oC Because low temperatu...

Page 11: ...S YOU HAVE PURCHASED A MASTER KEY Opening the Door Once the refrigerator has power you can now open the door using the factory set user code of 2244 Put in the code and after every correct digit a gre...

Page 12: ...rms and fridge function and allows you to download this information as a CSV file The controller will store this information for 18 months but we recommend that you download the data once a week espec...

Page 13: ...product The contents are not packed too closely together The contents do no touch the door wall or the bottom of the fridge and are stored away from the fan As with all refrigerators your Labcold phar...

Page 14: ...or you should contact your retailer in the first instance If you bought your fridge directly from Labcold you should contact our service team on 01256 705 580 or Email them at service labcold com When...

Page 15: ...www labcold com...