9
The internal light comes on when the door opens and go off when it is shut. It can be changed
so the light is either on or off all the time. To change the setting, access the menu in the usual
way by holding both the arrow icons/buttons and scroll through the menu until you get to the
Lt
screen or hold the light icon/button which will take you straight to the
Aut
screen.
Touch the tick icon/button as before to access this function and the screen will show
Aut
. This is the
default setting described above. To change the setting to the light being on all the time touch an
arrow icon/button until the screen displays
on
. Once selected, hold the tick icon/button until the
display reverts to
Lt
. To change the light operation to not coming on at all, select
OFF
instead and
again hold the tick icon/button until the screen reverts to
Lt.
You can either touch the mute icon/
button or just leave the controller alone and the screen will return to the normal temperature display.
10
If you do not fit batteries you will need to turn this alarm to
OFF
. To do so, hold the arrow
icons/buttons to enter the menu. Scroll through until you get to the screen that shows
bAt
and touch the tick icon/button. Touch the arrow icon/button to change the screen from
on
to
OFF
and hold the tick icon/button until the display reverts to
bAt
to confirm the change.
The battery symbol should then disappear from the temperature display screen and the alarm
won't sound.
There is a secondary temperature probe at the back of your refrigerator. It is a ‘wandering
probe’ you can move anywhere in the fridge, for example placing it in a simulated load. Like
probe
to1
the controller will alarm if probe
to2
detects that the temperature is lower than 2
o
C
or higher than 8oC where it is placed. The temperature detected by this probe will display in
normal operation after the
t1
screen. You can tell which one is which because a small
to1
or
to2
will appear above the temperature displayed.
Likewise you will be able to tell which probe is alarming as the high or low alarm will be
prefixed by
t1
or
t2
.
Internal Light
H/L
P
H/L
P
H/L
P
H/L
P
Battery Alarm
The battery alarm is enabled by default. This means if you put 4AA batteries (not supplied)
in your refrigerator the alarm will not sound unless the batteries are faulty or removed. For
extra security, the controller will alarm if the batteries are low or removed.
Labcold
strongly
recommends that you fit batteries to your
I
ntelli
C
old® pharmacy refrigerator for maximum
cold chain safety, however, should you decide not to you will need to disable the alarm, please
see
Turning the Battery Alarm Off
opposite
The alarm will show on the normal display as a battery symbol which will alternate with a
screen showing
n-b
. The alarm will sound when
n-b is
displayed. This alarm can be muted by
touching the mute icon/button while the
n-b
screen is displayed. The visual alarm will remain
and the screen
n-b
will still alternate with the temperature display. Unless you install batteries
the alarm will sound again after the time out interval is exceeded. (see page 8).
°
°t
1
C
H/L
P
H/L
P
Battery Alarm
Turning the Battery Alarm Off
H/L
P
H/L
P
H/L
P
Temperature Probe t02
H/L
P
°
°t
1
C
°
°t
C
2
Should you wish to disable it, enter the menu by holding the two arrow icons/buttons
simultaneously and scroll until you get to a screen that shows
t2
.
Touch the tick icon/button to enter the
t2
menu and the screen will display
on.
Touch the
arrow icon/button to reveal the
OFF
screen and hold the tick icon/button to confirm. The
display will then return to
t2
and then the normal temperature display.
Temperature detected
by probe
to1
Temperature detected
by probe
to2