10
If you do not fit batteries you will need to turn this alarm to
OFF
. To do so, hold both the arrow
icons/buttons to enter the menu. Scroll through until you get to the screen that shows
bAt
and touch the tick icon/button. Touch the arrow icon/button to change the screen from
on
to
OFF
and hold the tick icon/button until the display reverts to
bAt
to confirm the change.
There is a secondary temperature probe at the back of your refrigerator. It is a ‘wandering probe’
you can uncoil and move anywhere in the fridge, for example, placing it in a simulated load to
measure the thermal mass of the product. We recommend it is enabled as data collected by
this probe may be required for compliance and quality reasons etc.
Like probe
to1
the controller will alarm if it detects that the temperature is lower than 2
o
C or
higher than 8oC where it is placed. The temperature detected by this probe will display after
the
t1
screen. You can tell which one is which because a small
to1
or
to2
will appear above the
temperature displayed. Likewise you will be able to tell which probe is alarming as the high or
low alarm will be prefixed by
t1
or
t2
.
Turning the Battery Alarm Off
H/L
P
H/L
P
H/L
P
Temperature Probe t02
H/L
P
°
°t
1
C
°
°t
C
2
To enable the probe, enter the menu by holding the two arrow icons/buttons simultaneously
and scroll until you get to a screen that shows
t2
.
Touch the tick icon/button to enter the
t2
menu and the screen will display
OFF.
Touch the
arrow icon/button to reveal the
on
screen and hold the tick icon/button to confirm.
Temperature detected
by probe
to1
Temperature detected
by probe
to2