This refrigerator contains a fl ammable gas.
This refrigerator contains gas under pressure which may explode if heated.
The refrigerator MUST NOT be located in a room or area with a volume
less than 10m
3
.
Original instructions - Version 3 04.2020
WARNINGS
Servicing and repairs must be carried out by a suitably qualifi ed engineer.
Only engineers who have been trained in the safe handling and use of hydrocarbon
refrigerants can work on the refrigeration system.
If you suspect a leak contact the manufacturer or its service agents immediately.
Do not touch the refrigerant, it can cause severe eye irritation, redness, tearing, blurred
vision, and possible freeze burns. Contact with evaporating liquid may cause frostbite or
freezing of the skin.
If the power supply cord is damaged, it MUST be replaced by the manufacturer, its
service agent or similarly qualifi ed persons in order to avoid a hazard.
Do not use mechanical or other means to accelerate the defrosting process.
Keep clear of obstruction all ventilation openings in the appliance enclosure.
This refrigerator is climate class 4 and is designed to operate at +16°C to +32°C
ambient room temperature.
Do not damage the refrigerant circuit.
Maximum shelf loading 40 Kg/m
2
The noise level in normal operation is lower than 70 dB.
Do not use electrical appliances inside the refrigerator.
Do not store explosive substances such as aerosol cans with fl ammable
propellant in this appliance.
Labcold RLDF/G refrigerators are designed for the storing of pharmaceuticals,
medicines and laboratory products and are for professional use only by suitably
trained, competent persons.