Before Use

Thank you for purchasing a 

Labcold

 

I

ntelli

C

old® Pharmacy Refrigerator. Your fridge has been 

purposely designed and precision built for the safe storage of vaccines and other temperature 
sensitive pharmaceuticals.

Although great care has been taken in the manufacture of your refrigerator, damage can occur 
in transit. Before using for the fi rst time please remove all packaging and give your fridge a 
thorough visual inspection. If any damage is suspected, please contact your retailer immediately 
upon receipt of the product. Failure to do so may invalidate your warranty.

Setting Up

Place your fridge in the location in which it is to be used. 

 Do not place your fridge near a direct heat source or direct sunlight and leave at least 6cm all 

round the unit for adequate ventilation. Models designed to be wall mounted do not need extra 
space at the back because the cooling system is in the sides. Failure to do so may eff ect the 
performance of your refrigerator.

Clean your fridge with mild soapy water to remove any dust that may have accumulated during 
manufacture and dry thoroughly.

Ensure your fridge is level by adjusting the feet at the bottom of your fridge. 

                                                                   

For models RLDF/G0501A, 1010A and 1510A, insert shelves by placing clips in the racking and 
resting the shelf on top. When positioning the shelves, please remember to keep adequate room 
for air circulation. Failure to do so may cause hot spots in the fridge.

                                                                         

YOUR REFRIGERATOR SHOULD NOW BE LEFT FOR AT LEAST TWO HOURS BEFORE

CONNECTING TO POWER FOR THE FIRST TIME

2

A Closer Look

Alarm Battery Back Up

Remote Alarm and

Labcold accessories

sockets

Insert 4 AA long life alkaline batteries (not supplied) in the holder 
located on the right side of the unit 

 

Batteries will only power the controller not the fridge. 

Remote alarm and 

Labcold

 accessory sockets are located at the 

back of the fridge. If you have a 

Labcold

 external alarm kit this 

can be connected to the 2 pin connector on the left. The 4 pin 

connector socket is for the 

Labcold

 accessory kit.

The maximum rating for the connection is 1A 30Vdc. 

DO NOT CONNECT TO MAINS VOLTAGE

Door lock

Fan

Adjustable Feet

Adjustable Shelves

Maximum Load 

10kg

Led Light

Front Panel Display

Battery Holder

(side of the fridge)

Serial Plate

SD Card 

slot

Remote Alarm

and Accessory

Sockets

(back of fridge)

Summary of Contents for IntelliCold

Page 1: ...www labcold com IntelliCold Pharmacy Refrigerator Operating Instructions...

Page 2: ...irective and remain the user s responsibility to dispose of This refrigerator contains a flammable gas It MUST NOT be located in a room or area with a volume less than 10m3 This refrigerator contains...

Page 3: ...14 Digital Door Lock 15 Caring for your fridge 18 Loading 18 Cleaning 19 Defrosting 19 Trouble Shooting 20 Useful Information 22 IMPORTANT These operating instructions contain information essential to...

Page 4: ...during manufacture and dry thoroughly Ensure your fridge is level by adjusting the feet at the bottom of your fridge For models RLDF G0501A 1010A and 1510A insert shelves by placing clips in the rack...

Page 5: ...ber on the display will descend accordingly ALTHOUGH THE REFRIGERATOR WILL REACH 5oC AFTER AN HOUR APPROX YOU SHOULD LEAVETHE FRIDGE RUNNING FOR 24 HOURS BEFORE FILLINGWITHTEMPERATURE SENSITIVEVACCINE...

Page 6: ...asons for this such as extremely cold items being placed in the fridge The cause of the drop in temperature should be investigated immediately and the refrigerator should be monitored to check that it...

Page 7: ...uter and or print hard copy The information stored in the controller will cover the last 13 months and is not deleted from the fridges memory when downloaded It is recommended you perform a download o...

Page 8: ...n you have the correct day The display will then change to allow you to set the time Firstly you need to set the hour by using the buttons on the right of the display Press the tick button when you ha...

Page 9: ...s all the temperature data copied to it since the timer was first set on the refrigerator or the last time the data was downloaded from the controller The Labcold SD card can now be inserted into a SD...

Page 10: ...ntinually displays the first two digits of the model number ie 02 with no decimal point or the display does not change to an alarm state or show a temperature value after a few minutes there is a syst...

Page 11: ...around LEDS Red Blue Place the battery terminals on the correct contact points PP3 Battery NOT included Opening the door Once the refrigerator has power you can now open the door using the factory set...

Page 12: ...ion is paid to the loading of the fridge and it is regularly cleaned and defrosted If you are storing vaccines in your IntelliCold pharmacy refrigerator this should be recorded in accordance with reco...

Page 13: ...vaccines If you press the set button on the front it will show the temperature the fridge is set to Should you wish to change the set point please contact the Labcold service team at service labcold...

Page 14: ...idge is questioned by an official body Labcold are a UKAS calibration laboratory No 8898 and are therefore capable of providing calibration to ISO 17025 standards Why not contact our service team at s...

Page 15: ...www labcold com...

Reviews: