Plug in & switch on

Before plugging in your 

Labcold

 refrigerator for the first time make sure the power cable is free 

from damage and check that your voltage corresponds to that on the serial plate. 

It is recommended that you do not use extension cords or multiple socket adapters. Make sure 

the power cable is not bent, pinched or compressed and that it can be easily accessed.

Once plugged in, the word 

Labcold

 will scroll across the display followed by a two digit model 

number. Alarms will then show on the display and you may hear the audible alarm as the 

fridge completes its start up diagnostic checks. This is because the controller is factory set to 

alarm when it senses the temperature inside the refrigerator is above 8

o

C or below 2

o

C. The 

refrigerator is now working and the sensors are detecting the ambient temperature. Once the 

fridge has cooled sufficiently it will cease to alarm. You can press the ‘alarm mute’ button (see 

figure 1 below) to cancel the alarms.

The display will then show the temperature inside the fridge correct to one decimal place. This 

temperature is the actual temperature recorded by the probes inside the refrigerator and as 

the fridge cools to the set point (factory set at 5

o

C) so the number on the display will descend 

accordingly. 

ALTHOUGH THE REFRIGERATOR WILL REACH 5

o

C AFTER AN HOUR (APPROX) YOU SHOULD 

LEAVE THE FRIDGE RUNNING FOR 24 HOURS BEFORE FILLING WITH TEMPERATURE 

SENSITIVE VACCINES AND PHARMACEUTICALS.

                        

It is important that you reset the maximum/minimum temperature record after you have loaded 

the refrigerator for the first time (please see page 7). This will ensure that your daily temperature 

record is accurate when you start to use the fridge for storing medicines.

Figure 1 - Alarm mute button

Figure 2 - Alarm mute display

4

Front Panel Display

Opening for SD card

Press to view the 

set point of the 

fridge

Min/Max temperature 

display

Light on

Light off

Confirm button

Alarm Mute Button

High/Low Temperature Alarm

If the temperature in your refrigerator rises above 8

o

C an alarm will sound and the front 

panel will flash 

Hi

 and the current temperature inside the fridge. This information will display 

alternately. 

The cause of the rise in temperature should be investigated immediately. Usually, it is simply 

because the door has been open for a long time or the fridge has been restocked, however, it 

is recommended that you check the fridge to make sure that the temperature is returning to 

normal after the alarm has been activated.

Digital temperature 

display

The display on the front of the refrigerator is designed to constantly show the temperature 

inside the fridge correct to one decimal place. 

Summary of Contents for IntelliCold

Page 1: ...www labcold com IntelliCold Pharmacy Refrigerator Operating Instructions...

Page 2: ...irective and remain the user s responsibility to dispose of This refrigerator contains a flammable gas It MUST NOT be located in a room or area with a volume less than 10m3 This refrigerator contains...

Page 3: ...14 Digital Door Lock 15 Caring for your fridge 18 Loading 18 Cleaning 19 Defrosting 19 Trouble Shooting 20 Useful Information 22 IMPORTANT These operating instructions contain information essential to...

Page 4: ...during manufacture and dry thoroughly Ensure your fridge is level by adjusting the feet at the bottom of your fridge For models RLDF G0501A 1010A and 1510A insert shelves by placing clips in the rack...

Page 5: ...ber on the display will descend accordingly ALTHOUGH THE REFRIGERATOR WILL REACH 5oC AFTER AN HOUR APPROX YOU SHOULD LEAVETHE FRIDGE RUNNING FOR 24 HOURS BEFORE FILLINGWITHTEMPERATURE SENSITIVEVACCINE...

Page 6: ...asons for this such as extremely cold items being placed in the fridge The cause of the drop in temperature should be investigated immediately and the refrigerator should be monitored to check that it...

Page 7: ...uter and or print hard copy The information stored in the controller will cover the last 13 months and is not deleted from the fridges memory when downloaded It is recommended you perform a download o...

Page 8: ...n you have the correct day The display will then change to allow you to set the time Firstly you need to set the hour by using the buttons on the right of the display Press the tick button when you ha...

Page 9: ...s all the temperature data copied to it since the timer was first set on the refrigerator or the last time the data was downloaded from the controller The Labcold SD card can now be inserted into a SD...

Page 10: ...ntinually displays the first two digits of the model number ie 02 with no decimal point or the display does not change to an alarm state or show a temperature value after a few minutes there is a syst...

Page 11: ...around LEDS Red Blue Place the battery terminals on the correct contact points PP3 Battery NOT included Opening the door Once the refrigerator has power you can now open the door using the factory set...

Page 12: ...ion is paid to the loading of the fridge and it is regularly cleaned and defrosted If you are storing vaccines in your IntelliCold pharmacy refrigerator this should be recorded in accordance with reco...

Page 13: ...vaccines If you press the set button on the front it will show the temperature the fridge is set to Should you wish to change the set point please contact the Labcold service team at service labcold...

Page 14: ...idge is questioned by an official body Labcold are a UKAS calibration laboratory No 8898 and are therefore capable of providing calibration to ISO 17025 standards Why not contact our service team at s...

Page 15: ...www labcold com...

Reviews: