background image

7

Min/Max Temperature Recording

The DH Green Book Chapter 3 recommends that: ‘temperatures in the refrigerator must be 

monitored and recorded at least once each working day, and documented on a chart for 

recording temperatures’. 

Your 

Labcold 

pharmacy refrigerator records the maximum high and minimum low 

temperatures when connected to mains electricity (if batteries are fi tted, it will record when 

not connected). This information is stored until the fridge is reset.

This information can be accessed if you press the ‘Hi/Lo’ button and release. The display will 

show 

Hi

 and then the highest temperature recorded since the last reset, then 

Lo

 and display 

the lowest temperature. 

Once you have taken the temperatures for the day it is important that you reset the fridge. 

This is because the refrigerator will always display the high and low temperature since the last 

temperature reset. The refrigerator does not automatically reset.

To reset the minimum/maximum temperature press the ‘Hi/Lo’ button and hold. An alarm 

will sound and the current high and low temperatures will be displayed as above. Keeping 

your fi nger on the button, the display will then change to fl ash 

Hi Lo

 then revert to normal 

operation. This means you have successfully reset the minimum/maximum temperature 

display and the previous high and low temperatures have been deleted.

8

When you register the warranty of your 

Labcold

 refrigerator (UK customers only, available for 

purchase elsewhere), you will be sent an SD card. Although it looks like a card you may have 

in your camera, this SD card will only work with your 

Labcold

 fridge and using it for any other 

purpose may render it unable to have data from the fridge uploaded to it.

Your 

Labcold

 refrigerator automatically records performance information such as door 

openings, temperatures etc. The SD card allows you to extract this information and store it 

on your computer and/or print hard copy. The information stored in the controller will cover 

the last 13 months and is not deleted from the fridges’ memory when downloaded. It is 

recommended you perform a download of this data on a weekly basis and at least annually to 

prevent data being over written and the download failing. 

If you wish to store the data on the SD card and not your computer (which is preferable)  

we recommend you get a new card every year to prevent data overwriting. New cards are 

available at a small cost from 

Labcold

. Please e-mail 

sales@labcold.com

.

Only Labcold SD cards will work 

with Labcold Fridges

Using the SD Card

Requirements

Your 

Labcold 

RLDF/G10A series pharmacy refrigerator is fi tted with the unique 

I

ntelli

C

old® 

controller which stores temperature data for up to 13 months. To transfer this data to your PC 

you will require the following:

• 

A pre-programmed 

Labcold 

SD card supplied on return of warranty registration (UK 

customers only) or purchased separately. Only genuine 

Labcold 

SD cards will work with 

Labcold

 refrigerators.

• 

An SD card reader (not supplied) or port.

• 

CSV files will display on all PCs and Macs. 

• 

4 AA batteries (not supplied) fitted as shown on page 2 of the Operating Instructions. 

This will ensure that temperature logging and the internal timer will continue to operate 

during a power failure. Batteries must be changed yearly.

Please be aware that your 

Labcold 

SD card does not require any software to be downloaded 

to your computer and will not copy your data nor automatically download data to your 

computer.

YOU REQUIRE ONE LABCOLD SD CARD PER FRIDGE, OTHERWISE YOU RISK DATA BEING 

LOST OR OVER WRITTEN

We do not recommend that you leave the SD card in the refrigerator SD card 

slot as it may get broken or lost

 

 

Earlier models will reset to the set point ie 5⁰

C

. The correct high/low reading will replace 

this reading as the refrigerator is used, however, if the high and low reading remains at 5⁰

C

 

please contact the service team on 01256 705 580.

Min/Max Temperature Reset

Summary of Contents for IntelliCold

Page 1: ...www labcold com IntelliCold Pharmacy Refrigerator Operating Instructions...

Page 2: ...irective and remain the user s responsibility to dispose of This refrigerator contains a flammable gas It MUST NOT be located in a room or area with a volume less than 10m3 This refrigerator contains...

Page 3: ...14 Digital Door Lock 15 Caring for your fridge 18 Loading 18 Cleaning 19 Defrosting 19 Trouble Shooting 20 Useful Information 22 IMPORTANT These operating instructions contain information essential to...

Page 4: ...during manufacture and dry thoroughly Ensure your fridge is level by adjusting the feet at the bottom of your fridge For models RLDF G0501A 1010A and 1510A insert shelves by placing clips in the rack...

Page 5: ...ber on the display will descend accordingly ALTHOUGH THE REFRIGERATOR WILL REACH 5oC AFTER AN HOUR APPROX YOU SHOULD LEAVETHE FRIDGE RUNNING FOR 24 HOURS BEFORE FILLINGWITHTEMPERATURE SENSITIVEVACCINE...

Page 6: ...asons for this such as extremely cold items being placed in the fridge The cause of the drop in temperature should be investigated immediately and the refrigerator should be monitored to check that it...

Page 7: ...uter and or print hard copy The information stored in the controller will cover the last 13 months and is not deleted from the fridges memory when downloaded It is recommended you perform a download o...

Page 8: ...n you have the correct day The display will then change to allow you to set the time Firstly you need to set the hour by using the buttons on the right of the display Press the tick button when you ha...

Page 9: ...s all the temperature data copied to it since the timer was first set on the refrigerator or the last time the data was downloaded from the controller The Labcold SD card can now be inserted into a SD...

Page 10: ...ntinually displays the first two digits of the model number ie 02 with no decimal point or the display does not change to an alarm state or show a temperature value after a few minutes there is a syst...

Page 11: ...around LEDS Red Blue Place the battery terminals on the correct contact points PP3 Battery NOT included Opening the door Once the refrigerator has power you can now open the door using the factory set...

Page 12: ...ion is paid to the loading of the fridge and it is regularly cleaned and defrosted If you are storing vaccines in your IntelliCold pharmacy refrigerator this should be recorded in accordance with reco...

Page 13: ...vaccines If you press the set button on the front it will show the temperature the fridge is set to Should you wish to change the set point please contact the Labcold service team at service labcold...

Page 14: ...idge is questioned by an official body Labcold are a UKAS calibration laboratory No 8898 and are therefore capable of providing calibration to ISO 17025 standards Why not contact our service team at s...

Page 15: ...www labcold com...