15
Chapter 3: Getting Started
Connecting to the
Exhaust System (Optional
on XPert Filtered Balance
System)
The exhaust system should be installed by a
qualified HVAC contractor
.
The exhaust connection on the XPert Filtered
Balance Station has been designed to accept 6"
diameter ductwork as shown in Figure 3-2. The
exhaust connection ships uninstalled on the top and
needs to be fastened with the screws provided. See
Chapter 7 for ordering exhaust transition kits for the
XPert Filtered Balance System. Exhaust transition
kits aid in the removal of chemicals or applications
where a higher degree of powder and particulate
removal is required. Review Chapter 2 for exhaust
prerequisites and review Chapter 7 for ordering
blower exhaust equipment. See Figure 3-2 for the
XPert Station and Figure 3-3 for exhaust kit options
for the XPert Filtered Balance System.
Consult Labconco Customer Service should you
require help sizing your blower for the exhaust
volume and system static pressure loss.
To ensure compatibility, the selected exhaust duct
material should match the enclosure, procedures
and chemical applications.
!
WARNING: The weight of the exhaust
ductwork system must be supported
independently of the enclosure superstructure
or damage may occur.
!
XPert
Station
Summary of Contents for 3940200
Page 2: ......
Page 19: ...14 Chapter 3 Getting Started Figure 3 1 Filtered Enclosure Installation...
Page 73: ...68 Chapter 7 Accessorizing and Modifying your Filtered Enclosure...
Page 85: ...80 Appendix B Dimensions and Exhaust Options Figure B 1 XPert Filtered Balance System...
Page 86: ...81 Appendix B Dimensions and Exhaust Options Figure B 2 XPert Filtered Balance Station...
Page 90: ...85 Appendix C Filtered Enclosure Specifications C 1...
Page 91: ...86 Appendix C Filtered Enclosure Specifications C 2...
Page 92: ...87 Appendix C Filtered Enclosure Specifications C 3...
Page 93: ...88 Appendix C Filtered Enclosure Specifications C 4...
Page 100: ...95...
Page 101: ......