Chapter 6: Maintaining Your Filtered Enclosure
Annually
•
Replace the fluorescent lamps.
•
Have the enclosure validated by a qualified certification
technician. See Certification and Recertification in
Chapter
6
.
•
All monthly activities.
Decontamination
When used in conjunction with biohazards, the XPert Filtered
Balance System and XPert Filtered Balance Station should be
decontaminated with formaldehyde gas before:
•
maintenance work in contaminated areas
•
HEPA filter changes
•
moving the cabinet to a new location
•
changing research programs
•
after a gross spill of biohazardous material
The procedures for performing a gaseous decontamination are
thoroughly outlined in the U.S. Department of Health, Education
and Welfare booklet entitled
Formaldehyde Decontamination of
Laminar Flow Biological Safety Cabinets
, available from NIH,
Division of Safety, Bethesda, MD 20892, call 301-496-2801
Determination of when to Replace
HEPA Filters
The HEPA filters in the filtered enclosures gradually accumulate
airborne particulate matter and powders from the enclosure and
room. The rate of accumulation will depend upon the cleanliness
of the room air, the amount of time the enclosure is operating, and
the nature of work being done in the enclosure. In typical
installations and usage, the HEPA filters will last two to five years
before requiring replacement. Replace HEPA filters when face
velocity drops below the recommended 60-80 fpm velocity, and
the speed control is adjusted to full speed. Replace HEPA filters if
it fails the HEPA Filter Leak Test in Chapter 6.
Summary of Contents for 3940200
Page 2: ......
Page 19: ...14 Chapter 3 Getting Started Figure 3 1 Filtered Enclosure Installation...
Page 73: ...68 Chapter 7 Accessorizing and Modifying your Filtered Enclosure...
Page 85: ...80 Appendix B Dimensions and Exhaust Options Figure B 1 XPert Filtered Balance System...
Page 86: ...81 Appendix B Dimensions and Exhaust Options Figure B 2 XPert Filtered Balance Station...
Page 90: ...85 Appendix C Filtered Enclosure Specifications C 1...
Page 91: ...86 Appendix C Filtered Enclosure Specifications C 2...
Page 92: ...87 Appendix C Filtered Enclosure Specifications C 3...
Page 93: ...88 Appendix C Filtered Enclosure Specifications C 4...
Page 100: ...95...
Page 101: ......