37
Chapter 5: Using Your Filtered Enclosure and Appropriate Application
Hazardous Misapplications for
Odor Control Carbon Filters with
Volatile Chemicals
There is one scenario where the accessory carbon filter
misapplication would be a part of a hazardous condition. If the
user continues to operate the enclosure with any of the following
conditions present a potentially hazardous condition will exist:
1. The inlet concentration of vapors is greater than the TWA.
2. The carbon filter becomes saturated.
3. The ventilation of the room is insufficient to dilute the
exhaust of the enclosure to below the TWA for the
chemical.
When the inlet concentration is greater than the TWA, extra
measures must be taken to monitor the filter and number of room
air exchanges.
Chemical Carcinogen Use with
Odor Control Carbon Filters
Selected carcinogens may be used safely with Odor Control carbon
filters under the following restrictions.
The use of a vented fume hood or ventilated
enclosure with ducting to the outside is always
the preferred method when working with
carcinogens. The Odor Control carbon filters
should only be used, as a last resort when
venting to the outside is not an option.
The potential carcinogens are listed in the NIOSH Pocket Guide to
Chemical Hazards as “Ca.” Each potential carcinogen must have a
TWA of 1 or greater; have minimum respirator recommendation of
Escape GMFOV, and an odor threshold significantly lower than
the TWA for the chemical.
The inlet concentration or the evaporation rate of the chemical
must never exceed the TWA for the chemical.
!
Summary of Contents for 3980200
Page 2: ......
Page 13: ...8 Chapter 2 Prerequisites ...
Page 17: ...12 Chapter 3 Getting Started Figure 3 1 Filtered Enclosure Installation ...
Page 67: ...62 Chapter 7 Accessorizing and Modifying Your Filtered Enclosure ...
Page 82: ...77 Appendix C Filtered Enclosure Specifications C 1 ...
Page 83: ...78 Appendix C Filtered Enclosure Specifications C 2 ...
Page 84: ...79 Appendix C Filtered Enclosure Specifications C 3 ...
Page 85: ...80 Appendix C Filtered Enclosure Specifications C 4 ...
Page 92: ...87 ...
Page 93: ......