Labconco Glassware Washer Service/Technical Manual
13
To replace the conductivity sensor
1.
Turn off the electrical power to the washer. Turn off the circuit
breaker that services the washer.
2.
Remove the cabinet back panel.
3.
Unplug the sensor from the wiring harness.
4.
Unscrew the sensor from the sump housing.
5.
Reverse this procedure when installing a new sensor.
Note:
When replacing the sensor, use Teflon Tape on pipe threading.
Replacing the HEPA Filter
(FlaskScrubber® Vantage® only)
After an extended period of time, dirt and debris may block the air flow
through the HEPA filter.
1.
Open the left (side cabinet) door by pressing it in the center to
release the latch.
2.
Press down on the lock tab on the top of the HEPA filter and pull the
top of the filter outwards.
3.
Lift the filter out of the bottom retainer.
4.
Reverse this procedure to install the replacement HEPA Filter.