Labconco Glassware Washer Service/Technical Manual
23
Conductivity
reading “0”
Sensor not connected Reconnect sensor wires.
Sensor defective
Replace sensor.
Sensor dirty
Clean sensor.
Conductivity
Reading
fluctuates
Temp Reading
Stuck or
Fluctuating at
16ºC
Washing action cause
Bubbles
Poor Connection
between Sensor and
PC Board
None.
Check Connectors between Temp Sensor
and Sensor Wiring Harness.
Inspect integrity of Sensor Wiring
Harness.
Check Connectors between Sensor
Wiring Harness and Relay PC Board.
Check Connectors between Relay PC
Board and Ribbon Cable.
Inspect integrity of the Ribbon
Cable.
Check Connectors between Ribbon
Cable and PC Board.
Drain Fail Alarm
Drain Valve
inoperable
Listen for click when washer should
drain.
Check for mechanical restriction of
actuator arm on top of valve.
Check for electrical power to valve.
Check for damage to valve coil.
Drain Pump
inoperable
Check if pump fan turns when washer
should drain.
Check for mechanical restriction of
fan.
Check for electrical power to pump.
Drain hose not
routed
properly
Hose from drain pump must slope
upward and have no low spots or
horizontal runs under washer tank.
Make sure hose is attached to the
back of the washer and has a high
loop.
Hose must be attached to building
plumbing pipe ¾ inch ID or larger.